FURNACE DM
162 of 272
MAINTENANCE AND USE MANUAL
EN
(Conclusion of data entry for each program is, nevertheless, shown on the display by the
appearance of the word “OFF”).
The number “2” appears on the PHASE display (15).
D. Press
ENTER (28). On the control panel the following appears:
• the legend “tEmP” appears on the display (11) for one second.
• four numbers appear on the display (11) (these might even be 0000).
• the green STO-TEMP LED (24) come on.
• segment (2A) comes on.
Now enter the first temperature using the numerical keypad. Example: 250°C. Press keys 2 - 5 - 0.
E. Press ENTER (28) to save the temperature data. The following appears on the control panel:
• the green °C/min SLEW RATE LED (23) comes on.
The legend “rate” appears on the display (11) for one second. Now enter the increase rate
(in °C/min) that will be used to reach the just-set temperature. Do this via the numerical
keypad.
Example: 6°C/min. Press key number 6.
F. Press
ENTER (28) to save the rate data. The following then appears on the control panel:
• the green HR-MIN/TIME LED (22) comes on.
• the horizontal green segment (2B) comes on.
A second “time” appears on the display (11). Now insert the holding time via the numerical
keypad. Example: 20 minutes. Press keys 2 and 0.
G. Press ENTER (28) to save the data. Doing this completes PHASE 2 and brings the
operator to PHASE 3. The following then appears on the control panel:
• the green STO-TEMP LED (24) comes on again.
• the legend “tEmP” appears on the display (11).
• the green oblique segment (3A) comes on.
• the number “3” appears on the PHASE display (15).
H. Program PHASES 3 and 4 according to the procedure used in PHASE 2 as described above.
I. In
PHASE 5, after inserting the parameters as per the previous phases, the aspiration fan
cut-out temperature can be programmed (see section 7.4) by pressing the ENTER key (28):
• the legend “Fan” 300 appears for a fraction of a second on the display (11) (Aspiration Fan).
• the legend “U” appears on the display (15).
L. Press
ENTER (28):
• “t” appears on the display (15).
• the program duration, expressed in hours and minutes, appears on the display (11).
M. Press ENTER (28) for the last time to exit Data programming mode:
• the word OFF re-appears on the display (11).
To enter the data in the remaining programs, proceed as described in section 7.3.1 by keying in the
desired program number (0-9) during Program Selection (see 7.3.1, point B).
7.3.2
Entering data into remaining programmes
Summary of Contents for FURNACE DM 30
Page 1: ...FURNACE DM IT FR DE EN ES RU Preheating furnaces...
Page 2: ......
Page 3: ...IT Pag 5 FR Pag 51 DE Pag 95 EN Pag 139 RU Pag 225 ES Pag 183...
Page 4: ......
Page 48: ...48 di 272 MANUALE USO E MANUTENZIONE IT FURNACE DM A 6 TAVOLA PANNELLO COMANDI...
Page 92: ...92 de 272 MANUEL D UTILISATION ET D ENTRETIEN FR FURNACE DM A 6 DESSIN DU TABLEAU DE COMMANDE...
Page 136: ...136 von 272 HANDBUCH F R GEBRAUCH UND WARTUNG DE FURNACE DM A 6 SCHALTPULT...
Page 180: ...FURNACE DM 180 of 272 MAINTENANCE AND USE MANUAL EN A 6 CONTROL PANEL...
Page 224: ...FURNACE DM 224 de 272 MANUAL DE USO Y MANTENIMIENTO ES A 6 TABLA PANEL DE CONTROL...
Page 228: ...RU 228 272 FURNACE DM 1 2...
Page 230: ...RU 230 272 FURNACE DM Zhermack 2 4 2 3 Zhermack 6 3 6 3 7 1 2 3...
Page 231: ...RU 231 272 FURNACE DM Zhermack Zhermack 12 8 1 2...
Page 240: ...RU 240 272 FURNACE DM 5 1 5 2 Zhermack DM 4 1 2 5 Zhermack Zhermack...
Page 241: ...RU 241 272 FURNACE DM 1 2 5 3 a 2 b c 3 4 1 1...
Page 242: ...RU 242 272 FURNACE DM 5 4 2012 19 UE 5 4 1 5 4 2 3 180 m CR2032 FURNACE DM...
Page 243: ...RU 243 272 FURNACE DM 6 1 6 3 6 3 6 2 200 2 4 2000 5 C 40 C 80...
Page 244: ...RU 244 272 FURNACE DM 6 3 1 DM 40 7 15...
Page 245: ...RU 245 272 FURNACE DM DM1 DM1G 6 4 16 32 DM1 DM1G 7 14...
Page 246: ...RU 246 272 FURNACE DM 3 6 3 6 3 1 6 4 6 5 3 4 4 B DM 6 4 1 DM 30 40 50 DM4R 1 1 2 2...
Page 258: ...RU 258 272 FURNACE DM DM 4R DM 7 15 DM 4R 7 7 0 2 N 8 8 ENTER 7 16 DM...
Page 259: ...RU 259 272 FURNACE DM 3 7 1 D...
Page 260: ...RU 260 272 FURNACE DM 8 1 PULIZIA DELL APPARECCHIATURA 8 2 8 1 2...
Page 262: ...RU 262 272 FURNACE DM 9 9 1 9 2 1 2 1 2 1 1 1 1 tEmP 1 7 2 B 2 1 2 11 Door 1 1...
Page 263: ...RU 263 272 FURNACE DM A 1 A 2 A 4 A 5 A 6 A 3...
Page 266: ...RU 266 272 FURNACE DM A 3...
Page 268: ...RU 268 272 FURNACE DM A 6...
Page 269: ...FURNACE DM...
Page 270: ...FURNACE DM...
Page 271: ......
Page 272: ...XM0320000 16 10...