background image

19

hurrimix

The reference figures are present on pages 3, 4, 5 and 6 of this manual and in the product Quick Guide.

 

EN

 

7.2 

USE OF ThE MIXER WITh ALGINATE

Do  not  activate  the  device  without  material  inside  the 
container.

hurrimix

  mixes  the  most  common  alginates; 

recommended quantity 60g, max 80g. Do not exceed the 
maximum permitted quantity to avoid malfunction.

The device guarantees the best results by using Zhermack 
alginates.

Once  it  has  been  correctly  installed  and  put  into  service,  proceed 
as follows:

Fig. 5: Press the program button (fig. A, 8) to scroll and select the 
desired program in the Alginates section (fig.5, A). Mixing time can 
be customised, by pressing buttons “+” (fig.A, 9) and “-” (fig.A,10). 
Keep the button pressed for a fast scroll.
Fig. 6: Dose the alginate powder with a measuring cup as suggested 
by  the  Manufacturer  and  pour  it  in  the  appropriate  container  of 
alginate mixing.

Comply with the dosage indicated by the Manufacturer. 

Fig.  7:  Add  the  correct  quantity  of  water,  using  the  bottle  with 
graduated measuring cup (provided with).
Fig. 8: Slowly rotate the container with the mixture for a moment.
Fig.  9:  Close  the  mixing  container  by  pressing  the  cover  and 
then turning it clockwise until it locks in place.

It is important to carry out the pre-mixing manually, so 
that the powder can be entirely absorbed by the water.

To  prevent  leakage  of  material,  and  malfunctions, 

ensure that the cover of the container is closed properly

.

The  mixing  container  is  liable  to  wear  and  should  be 
replaced after a period of time.

Fig.  10:  Place  the  container  inside  the  equipment  and  close  the 
door.
Fig.  11:  Once  the  program  has  been  selected  (Fig.5,  A),  press  the 
START button to start a mixing cycle. 

hurrimix

  mixing  accelerates  the  setting  time  of  the 

mixture; use cold water to limit the phenomenon.

The  device  will  stop  automatically  when  the  selected 
mixing cycle finishes.

Fig. 12: Once the cycle has finished, the mixture is ready to be used: 
Use  the  provided  spatula  to  collect  easily  the  mixture  from  the 
alginate container. 

We strongly advise never mixing alginate then stone in 
the same container as it may contaminate & impede the 
setting.

It  is  recommended  to  remove  the  mixture  from  the 
container as soon as possible.

7.3 

USE OF ThE MIXER WITh PLASTER

Do  not  activate  the  device  without  material  inside  the 
container.

hurrimix

  mixes  Zhermack  Elite  Dental  Stones  only; 

recommended  quantity  80g,  max  100g.  Do  not  exceed 
the maximum permitted quantity to avoid malfunction.   

Once  it  has  been  correctly  installed  and  put  into  service,  proceed 
as follows:
Fig. 13: Press the program button (fig. A, 8) to scroll and select the 
desired  program  in  the  Dental  Stones  section  (fig.13,  B).  Mixing 
time can be customised, by pressing buttons “+” (fig.A, 9) and “-” 
(fig.A,10).

6.6 

ZhERMACK PRESET PROGRAMS 

hurrimix 

software has 20 mixing programs, divided as follows:

- alginate section: 10 programs, 7 of which are preset (specific for 
Zhermack products) and 3 can be managed by the user. 
-  plaster  section:  10  programs,  7  of  which  are  preset  (specific  for 
Zhermack products) and 3 can be managed by the user.

All  programs  can  be  modified  by  the  user:  By  pressing 
buttons “+” and “-”, the mixing time can be increased or 
decreased  (from  a  minimum  of  2  to  a  maximum  of  30 
seconds). 
 

   

                

Alginates 

Time Mixing

Hydrocolor 5 

6 sec

Hydrogum 5 

6 sec

Hydrogum 

10 sec

Orthoprint 

6 sec

Tropicalgin 

8 sec

Phase Plus 

8 sec

Neocolloid 

10 sec 

Prog. 1 

6 sec

Prog. 2 

8 sec

Prog. 3 

10 sec

Dental Stones 

Time Mixing

Elite Rock 

6 sec

Elite Master 

6 sec

Elite Model 

8 sec

Elite Model Fast 

6 sec

Elite Stone 

6 sec

Elite Ortho 

6 sec

Elite Arti 

8 sec

Prog. 1 

6 sec

Prog. 2 

8 sec

Prog. 3 

10 sec

To reset the default programs, press and hold button 8 (fig. 
A)  and  at  the  same  time,  turn  the  device  on  by  pressing 
button 7 (fig. A). 

6.7 

DECLARED USE

hurrimix

 mixer has been designed to mix alginates and plasters for 

the dental industry.

6.8 

NON-DECLARED/PROhIBITED USE

Using the equipment for any purpose not considered in 
paragraph regarding “Declared use” shall be considered 
improper and thus forbidden. 
The improper use of the equipment renders the warranty 
null and void and Zhermack Spa declines all liability for 
any damage to objects and injury to operators or third 
parties.

The  main  reasons  for  which  the  warranty  is  rendered  null  and 
void are given in paragraph 2.2 and in the “Warranty Certificate” 
attached to this Manual.

ChAPTER 7: USING ThE EqUIPMENT

7.1 

GENERAL WARNINGS

For total certainty of safe operation, scrupulously follow 
the work phases given below.

For  greater  clarity,  the  numbers  stated  in  the  figures 
positioned  near  to  the  paragraph  boxes  correspond  to 
the  numbering  of  the  data  in  Chapter  3,  regarding  the 
“Description of the equipment”.

To  avoid  the  occurrence  of  any  risks,  pay  attention  to 
avoid  pouring  water  during  the  preparation  on  the 
electrical voltage parts.

To  avoid  any  interference,  the  equipment  has  been 
projected to perform a mixing cycle every 3 minutes.

Summary of Contents for Hurrimix

Page 1: ...t e R o c k E l i t e M a s t e r E l i t e M o d e l E l i t e S t o n e E l i t e M o d e l F a s t E l i t e A r t i P r o g 1 P r o g 2 P r o g 3 P h a s e P l u s H y d r o c o l o r 5 A l g i n...

Page 2: ......

Page 3: ...hurrimix 3 4 5 6 1 2 8 7 9 10 11 3 13 14 15 12 16 A C A B C D BEYOND INNOVATION ZHERMACK SpA Badia Polesine RO Italy C303230 HURRIMIX 230V SN xxxxxxxx 230V 50 60Hz 400W B D h u rr im ix E F...

Page 4: ...tus Elite Ortho Elite Rock Elite Master Elite Model Elite Stone Elite Model Fast Elite Arti Prog 1 Prog 2 Prog 3 Phase Plus Hydrocolor 5 Alginates Hydrogum 5 Hydrogum Tropicalgin Mixing Time sec 06 Or...

Page 5: ...o p r in t N e o c o ll o id P r o g 1 P r o g 2 P r o g 3 10 7 2 3 sec 8 Dental Stones Status Elite Ortho Elite Rock Elite Master Elite Model Elite Stone Elite Model Fast Elite Arti Prog 1 Prog 2 Pro...

Page 6: ...g and unpacking 5 3 Transport and handling 5 4 Disposal Dismantling 6 INSTALLATION AND COMMISSIONING 6 1 Warnings and precautions 6 2 Accepted environmental conditions 6 3 Space necessary for use 6 3...

Page 7: ...der to guarantee your safety and that of others and to prevent damage to the equipment Any operations carried out without compliance to instructions or tampering with equipment parts which can cause d...

Page 8: ...hnologically advanced work instruments in away that the user feels that they are made to measure his everyday working requirements Zhermack S p A uses state of the art instruments in the design field...

Page 9: ...keep the original packaging so that if necessary it can be re used for transport and handling the equipment thus preventing accidental damage due to the use of unsuitable packaging see paragraph 5 3...

Page 10: ...Stones only recommended quantity 80g max 100g Do not exceed the maximum permitted quantity to avoid malfunction Once it has been correctly installed and put into service proceed as follows Fig 13 Pre...

Page 11: ...ASTER CONTAINER Rinse the container and cover before the mixture hardens Gently clean the surfaces with non alcohol based products Do not use inflammable corrosive alkaline or toxic liquids to clean t...

Page 12: ...g from poor maintenance incorrect power supply negligence inexperience and other causes not imputable to the Manufacturer and parts subject to normal wear are not covered by the warranty The warranty...

Page 13: ...22 hurrimix The reference figures are present on pages 3 4 5 and 6 of this manual and in the product Quick Guide EN 10 3 TECHNICAL INTERVENTIONS SHEET Date Description of the interv Signature...

Page 14: ......

Page 15: ...XM2501510 Rev 05 12 ww w tec nomedia biz...

Reviews: