Be absolutely certain that only the specified fuses
as above are used. The use of fuse with different
specifications may place the operator, the system
& measuring instrument in danger. The use of
repaired fuses or short-circuting of the fuse holder
is prohibited.
Replace the housing base and press until it snaps audibly
into place.
Dispose of the dead battery in an environmentally sound
fashion.
16.2 Fuses
The 16 A fuse interrupts the 10 A current measuring range,
and the 1.6 A fuse the mA current measuring ranges. All
other measuring ranges continue to function.
If a fuse blows, eliminate the cause of the overload before
placing the instrument back into operation!
Replacement of Fuses
Open the instrument as described under battery replacement.
Remove the defective fuse with the help of, for example, a
test probe, and replace it with a new fuse.
Make certain that the new fuse makes good contact.
The following fuses may be used :
* for current measuring ranges up to 4mm mA :
type FF 1.6 A / 600 V~ ; 6.3 mm x 32 mm
* for the 10 A measuring range :
type FF 16 A / 600 V~ ; 6.3 mm x 32 mm