Test conditions:
EN 61010-1
Rated impulse withstand voltage
4000 V
Surge category
III
Pollution level
2
Rated insulation voltage Ui
300 V
Operating time
100 %
Galvanic insulation
Supply voltage
– Measurement input DC 3820 V
Ethernet Control voltage
–Measurement input
DC 500V
EMC-tests:
EN 61326-1
emitted interference
EN 61000-6-3
Burst
EN 61000-4-4 +/-4 kV
Pulse 5/50 ns, f = 5 kHz, t = 15 ms, T = 300 ms
SURGE
IEC 61000-4-5 +/-4 kV
Impulse 1,2/50 µs (8/20 µs)
discharge of static electricity
IEC 61000-4-2, +/-4 kV contact, +/- 8kV air
Environmental conditions:
Ambient temperature range
-20 °C ... +65 °C
Storage temperature range
-
20 °C … +70 °C
Altitude
Up to 2000 m
Climatic conditions
5
– 85 % rel. humidity, no condensation
External wiring temperature range
-
5 °C … +70 °C
Vibration resistance EN 60068-2-6
2…25 Hz ±1,6 mm, 25 ... 150 Hz 5 g
Housing:
Type V8, distribution board
Dimensions (W x H x D)
140 x 90 x 58 mm
Front-to-back size / Width 55 mm / 8 TE
Wiring connection single strand
each 1 x 1.5 mm
2
Finely stranded with wire end ferrule
each 1 x 1.0 mm
2
Starting torque of the terminal screw
0.5 Nm (3.6 lb.in)
Protection class housing / terminals
IP 30 / IP20
Mounting position
Arbitrary
Mounting
Snap-on fastening standard rail
35 mm acc EN 60715 or
Fasten with screws (with 2 additional bars)
Weight:
Approx. 350 g