background image

 

DCO-1318 

 

Effective Date: 01-SEP-2021

 

LBL 6028 Rev. G 

Page 3 of 23 

 

WARNING:

 DO NOT use sterile devices after the stated expiration date.

 

  

 

WARNING:

 Continually check all handpieces and 

attachments for overheating.  If overheating is sensed, 

immediately discontinue use and return equipment for service.  

Overheating can lead to patient injury (burn). Refer to Tricera Handpiece 

Instructions for Use LBL 6025/LBL 6043 for additional warnings and 

precautions for the Tricera Handpieces (All Models).

 

WARNING:

 The service intervals are required to keep the handpieces 

at their optimum operating performance.  Failure to follow the 

recommended PM schedule can cause devices to overheat which may 

lead to patient injury (burn). Refer to Tricera Handpiece Instructions for 

Use LBL 6025/LBL 6043 for additional warnings and precautions for the 

use of the Tricera Handpieces (All Models). 

WARNING: 

DANGER: DO NOT operate the Tricera System in close 

proximity to volatile solvents such as methanol or alcohol, or in the 

presence of flammable anesthetics, as explosion may occur. 

WARNING: 

DO NOT connect the Tricera FMA to another fluid 

management system, shaver handpiece, or controller.  Use with another 

fluid management system, shaver handpiece or controller may result in 

failure of the device to operate or lead to patient or physician injury.   

WARNING:

 Equipment grounding is vital to ensure safe operation.  Plug 

the power cord into a properly earthed mains supply outlet whose 

voltage and frequency characteristics are compatible with those listed 

on the unit or in this manual.  DO NOT use plug adapters or extension 

cords; such devices defeat the safety ground and could cause injury. 

Fluid Management Warnings

 

WARNING: 

Place the pump at the same height or lower than the joint to 

ensure accurate pressure readings. 

WARNING: 

Perform the endoscope calibration at the same height as 

the joint to ensure accurate pressure readings.  Ensure that the 

endoscope stopcock connected to the inflow tubing is completely open 

during the calibration procedure and that any other stopcock is closed.   

WARNING: 

Using fluid to distend the joint carries the possibility of fluid 

extravasation into the surrounding tissue. Select the optimal pressure 

based on the patient profile, including but not limited to: blood pressure, 

height, weight, age, and tissue quality.  Recommended pressure 

settings are included in this manual; however, these are only 

suggestions, and each surgery and each patient may require different 

parameters. 

WARNING: 

The fluid pump is only intended for use with flexible 

containers for the inflow fluid.  Do not use glass containers as they 

might implode due to the vacuum being generated inside of the 

container.

 

WARNING: 

For Veriflow inflow only, DO NOT partially open or close the 

stopcocks on the inflow sheath as this may lead to overpressure or 

underpressure within the joint.  The inflow stopcock should be 

completely open or completely closed. 

WARNING: 

Keep the pump dry. If liquid has accidentally leaked into the 

pump from the Fluid Management Accessories cassette, change the 

cassette.

  

WARNING: 

Failure of the system may result in an unintended increase 

or decrease in flow and/or pressure.  Carefully monitor the joint when 

using the system. 

Known Risks Associated with use of Electrosurgical Devices:

 

1.

 

EMC issues – interference causes device failure, 

interference causes other devices to fail, RF interferes with 

pacemaker, defibrillator 

2.

 

Electrical safety issues – shock, burn – device/controller 

overheats, incorrect power source used, fluid enters the 

controller, use of incorrect power source, arcing.

  

3.

 

Explosion/fire if operated near volatile solvents 

4.

 

Tissue damaged during coagulation/resection 

Shaver Blade/Burr and RF Probe Warnings 
WARNING: 

DO NOT operate the Tricera Devices without clear 

visualization.  The device should be in the field of view while it is 

operating.  If visualization is lost at any point during the procedure, the 

device’s function must be stopped immediately. 

WARNING: 

Do NOT touch the device tip at any time during the 

application of motor or RF power.

 

WARNING: 

Do NOT insert or withdraw the device during application of 

power.

 

WARNING: 

Do NOT activate the device while in contact with metal 

objects and/or instruments as unintended tissue or device damage can 

occur. 

WARNING: 

During extended periods of activation the shaft of the 

device may become hot.  Pause occasionally to keep the shaft cool. 

WARNING: 

Ensure that adequate inflow and outflow of conductive 

irrigant is maintained at all times during RF probe use or else irrigant 

may overheat and damage surrounding tissue. 

6. 

PRECAUTIONS 

A precaution contains instructions for any special care to be exercised 

by the practitioner for the safe and effective use of the equipment.  

Failure to observe the information or procedures presented in a 

Precaution may result in damage to the equipment. 

Tricera System General Precautions 

 

Federal (US) law restricts this device to sale by or on the order of 

a physician. 

 

DO NOT expose the Tricera Controller to moisture, operate in wet 

areas, or place liquids on or above the controller. 

 

Tricera handpieces are factory sealed.  DO NOT disassemble or 

lubricate, as this may void the warranty.  There are no user-

serviceable parts inside. 

 

Use only associated Tricera attachments and accessories as 

defined within the descriptions of each attachment. 

 

After each use, thoroughly clean the controller, footswitch, and 

handpieces. 

 

DO NOT use the shaver blades/burrs and RF probes or the Fluid 

Management Accessories if the sterile barrier or sterility is 

compromised prior to or during the procedure.  Failure to maintain 

sterile technique in the operating room could result in the transfer 

of microorganisms into the body tissues.   

 

The shaver blades/burrs and RF probes and Fluid Management 

Accessories are intended for single patient use only.  Discard 

these disposable devices after each patient use. 

Summary of Contents for Tricera

Page 1: ...15 Appendix C Abbreviations 15 Appendix D Tones 15 Appendix E Troubleshooting 16 Appendix F Essential Performance Power Curve 18 Appendix G EMC Tables 19 Appendix H User Profiles 21 1 SYMBOLS USED ON...

Page 2: ...on during diagnostic and operative arthroscopic procedures The Shaver Blade RF Probe of the Tricera Arthroscopic System provides abrasion resection debridement and removal of bone through its shaver b...

Page 3: ...hin the joint The inflow stopcock should be completely open or completely closed WARNING Keep the pump dry If liquid has accidentally leaked into the pump from the Fluid Management Accessories cassett...

Page 4: ...ble devices such as cardiac pacemakers If the patient has an implantable cardioverter defibrillator ICD contact the ICD manufacturer for instructions before performing ablation coagulation or other el...

Page 5: ...ay result in damage to the tip of the devices 7 ENVIRONMENTAL PROTECTION Observe all national waste management regulations DO NOT dispose of Waste Electrical and Electronic Equipment WEEE as unsorted...

Page 6: ...6 Receptacle reserved for future functionality 7 Power ON OFF Switch and Power ON LED 8 Pressure Sensors Figure 2 Controller Rear 9 Equipotential Lug 10 Power Entry Module 11 Fuse Drawer 12 USB Acces...

Page 7: ...nt 6 Inflow to Patient 7 Inflow Pressure Diaphragm 8 Inflow Peristaltic Pump Tube 1 TPE Pump Tubing 2 Barb Lock 3 PVC Waste Tuing 4 Outflow Lumen 5 Barb Connector 6 Cassette Magnet Figure 5 Veriflow T...

Page 8: ...Date 01 SEP 2021 LBL 6028 Rev G Page 8 of 23 Figure 7 Tricera Handpiece LCD and Buttons 1 Activate 2 Coagulation 3 Joystick 4 Mode 9 5 Shaver Blades 1 Coupler 2 Hub 3 Shaft 4 Tip Figure 8 Tricera Devi...

Page 9: ...x H for user profile details 5 The Zimmer Biomet logo and self test progress will appear on the screen followed by the images shown in Figure 11 and 12 Figure 11 Figure 12 6 Open the FMA Cassette pack...

Page 10: ...e end of the Tricera handpiece as shown Figure 18 NOTE steps 14 17 apply to Veriflow FMA set R 10003 only 14 Select the joint for the procedure by touching the joint selection touchscreen or using the...

Page 11: ...at any time during the procedure 2 The Tricera Controller will set the following default pressure set point based on the initial joint selection Knee 45mmHg Shoulder 50mmHg Hip 50mmHg Small Joint 35m...

Page 12: ...e the potential of blood and debris drying prior to cleaning 3 After use Tricera Devices may be a potential biohazard and should be handled in accordance with acceptable medical practice and applicabl...

Page 13: ...5 4 Allow 0 Suction YES and NO When this setting is set to Yes the suction level for Oscillate Bone Cutting Bone Burring Forward and Reverse all non RF modes shall be adjustable between 0 and 100 When...

Page 14: ...of merchantability and fitness for a particular purpose and all other obligations and liabilities on the part of Zimmer Biomet Zimmer Biomet neither assumes nor authorizes any other person to assume f...

Page 15: ...Pressure 878 to 1082 cmH2O 86 to 106 kPa IV Transport and Storage Requirements Temperature 18 C to 60 C Relative Humidity 15 to 85 non condensing Atmospheric Pressure 510 to 1082 cmH2O 50 to 106 kPa A...

Page 16: ...tomer service E11 E16 Inflow turns off High pressure detected on the inflow line Foreign object between diaphragm and pressure sensors Controller malfunction Remove any obstructions from inflow line o...

Page 17: ...t the footswitch is plugged into the gray port on the Controller Ensure the Handpiece is securely plugged into the connector Replace Tricera Device N A N A N A F12 Handpiece Cannot Position Window Pro...

Page 18: ...ntentional activation of RF output no unintentional activation of pump motors and correct pressure indication within one indicator bar II Power Curve Full and Half Output Power versus Load Resistance...

Page 19: ...ootswitch 10 ft Hospital Grade Power Cord Table 1 Electromagnetic Emissions Statement Guidance and Manufacturer s Declaration Electromagnetic Emissions The Tricera System is intended for use in the el...

Page 20: ...lity should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 4...

Page 21: ...3 79 3 79 7 27 100 12 0 12 0 23 0 For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to...

Page 22: ...3000 6000 Forward RPM Reverse Default 1000 6000 6000 Reverse RPM Bone Cutting Default 3000 15000 15000 Bone Cutting RPM Bone Burring Default 3000 15000 15000 Bone Burring RPM RF Ablation Default 1 10...

Page 23: ...handpiece shall continue running when the button is released When a button is pressed again the motor shall stop running In RF modes or when Handpiece Button Latch non RF modes is Off activation shall...

Reviews: