Cleaning
Disinfection
10
Page 17
-Before starting any maintenance and cleaning measures the device must always be
turn off at the main switch and unplugged.
- Make sure that when cleaning and disinfecting the labels of the device (such as
warnings, labels of control devices, identification plate) are not damaged.
- Make sure, when cleaning and disinfecting, that no liquids penetrate the device. Do
not use sprays.
- If during cleaning or disinfecting liquid penetrates the device, please put the unit out
of service, protect it from getting used again and contact your service representative.
- The device and its applied part are not considered critical in relation to hygiene when
used on non-injured and healthy skin.
Housing / Accessories Cleaning: In the event of visible contamination, the housing, accessories and all
cables can be cleaned using commercially available soft alcohol-free plastic cleaners.
Wipe the surface until the dirt is removed, using a soft cloth soaked according to the
specifications of the manufacturer of the cleaning agent but not dripping.
Disinfection: We recommend that disinfection to be carried out at least once a week,
as well as in the event of evidence of possible contamination. Consult with your health
professional when doing so. Always perform cleaning prior to disinfection.
Housing, cables and accessories can be disinfected with disinfectant wipes. To do so,
use a commercially available alcohol-free disinfectant for metal and plastic, which is
bactericidal, virucidal and fungicidal. Observe the application instructions of the
manufacturer. Wipe all surfaces with a non-dripping soft cloth, soaked according to the
specifications of the manufacturer of the cleaning agent, or with pre-soaked
disinfectant wipes.
Please note, where appropriate, specifications for drying or post-cleaning.
Container for defrosted
water
Cleaning: The container for defrosted water (7) should be cleaned after each draining
of condensed water. Proceed as indicated under "Housing / Accessories".
Disinfection: Proceed as indicated under "Housing / Accessories".
Attention !
The device may only be operated with the container for defrosted water installed.
Air filter
The air filter (9) should be cleaned regularly, however, not later than after 200
operating hours (maintenance message will appear in the display).
For this, the air filter should be cleaned from the outside with a standard household
vacuum cleaner.
Note:
Use the device only in hygienic environment.