Functional Test
14
Page 34
enPuls
Pro
performs a self-test when switched on checking all the internal
components.
In the event of a fault, an error message appears.
In addition, a further function test can be performed as described below.
This test should be performed monthly or in case of doubt about the proper
function of the device.
Note:
Before performing the function test, check that the handpiece and footswitch
are properly connected to the device.
Check that the mains cable is properly connected to the device and that it is
connected to the mains.
Function test
Switch the device on.
Briefly activate the footswitch - the fans and generator should start
immediately, whereby the radial pulse generator must operate at the frequency
shown on the display (5 Hz is the default value).
Note:
After completing the function tests, switch off the enPulsPro.
If a therapy is to be carried out directly afterwards, set the desired treatment
parameters and proceed as described in Chapter 8.