Error Messages
Troubleshooting
Disposal
16
Page 37
Applicator not found
In the status bar, the message "No applicator found" appears.
Possible cause
Handpiece not or improperly connected.
Remedy for cause
Make sure that the handpiece is properly connected. The plug must be fully
inserted.
Device malfunction
No response to the main switch / display remains dark
Possible cause 1
Mains power connection
Remedy for cause 1
Check whether the mains cable is properly plugged into the electrical outlet
and that the device connector is firmly inserted into the socket of the device.
Check the mains cable for damage. Replace it in the event of visible damage.
Check the power supply and mains socket.
Possible cause 2
Fuse
Remedy for cause 2
In the power input socket on the device there are micro-fuses that disconnect
the power supply in the event of an electrical problem. Open the cover and
check the fuses. If necessary, replace the faulty fuse.
Only replace the fuse with a replacement that has the same name / is an exact
equivalent. Perform a complete check of the power supply for possible faults
beforehand.
If the fault persists, contact the service representative / customer service
immediately.