OPERATING PROCEDURE
12
•
STERILE PERSON (A):
Assemble the funnel on the housing.
•
NON-STERILE PERSON (B)
: Insert battery into housing through funnel opening.
Push until fully inserted, and remove funnel.
A
B
A
B
A
B
A
Funnel
A
•
STERILE PERSON (A)
: Close lid. Position the locking system to
to lock the housing.
A
A
Locked
Unlocked
Once a “click” has been heard, the battery pack is ready to use.
TO REMOVE BATTERY FROM ATK HOUSING
After the procedure is completed, the
NON-STERILE PERSON (B)
disassembles the Aseptic Transfer Kit (ATK) by removing the
battery from the ATK housing. The instructions from the prior section should be followed with the steps in reverse manner,
without using the funnel.
WARNING
: The
Zimmer®
UNIVERSAL battery (REF 89-8510-440-20) for Aseptic Transfer Kit (ATK)
must not be cleaned in an
automated washer or sterilized.