DESCRIPTION AND TECHNICAL DATA
5
Environmental conditions during operation
Ambient temperature
10°C to 38°C (50°F to 100°F)
Relative humidity
30% to 60%
Altitude
≤2,000 m – Not intended for high altitude environments
NOTE
: There are no special requirements for transport or storage. Transport or store in an environment that limits exposure
to dust, moisture and temperature extremes.
CAUTIONS
:
•
Use only with
Zimmer®
UNIVERSAL Battery / Electric handpieces and
Zimmer®
UNIVERSAL sterilizable power cord.
•
Use exclusively a certified main power cord integrating earth connection.
•
Do not use the electric power supply if it has been damaged or immersed.
•
Ensure the sterilizable power cord does not create a trip hazard.
•
Check that the voltage of the mains power supply is 100-240 volts.
•
Before unplugging power cords, switch the handpiece and power supply off.
•
Never block the air vents.
•
Do not immerse power supply or power cords.
•
The ON/OFF switch on this unit will not completely shut off all power from the AC outlet. Since the power cord serves as the
main disconnect device for the unit, you will need to unplug it from the AC outlet to shut down all power. Therefore, make
sure the unit has been installed so that the main power cord does not create a trip hazard and can be easily unplugged
from the AC outlet if needed
COMPACT BATTERY CHARGER
REF 89-8510-421-00
•
Electric battery charger
•
Compatible with
Zimmer®
UNIVERSAL battery
for Aseptic Transfer Kit and
Zimmer
® UNIVERSAL
sterilizable battery
•
Able to charge four batteries simultaneously
•
Indicator light shows charge status of battery
Technical data
Input
100-240V~, 50-60Hz, 2-1A
Output
4 x 18V DC, 1.5A
Dimensions
321 mm x 206 mm x 94 mm
Weight
2.7 kg approximately (without battery)
Type of protection against electric shock
Class I
Environmental conditions during operation
Ambient temperature
5°C to 25°C (41°F to 77°F)
Do not operate in patient area
Relative humidity
30% to 60%
NOTE
: There are no special requirements for transport or storage. Transport or store in an environment that limits exposure
to dust, moisture and temperature extremes.
ENGLISH
Summary of Contents for Universal Power System
Page 1: ...Zimmer UNIVERSAL Power System Instruction Manual Battery Electric ...
Page 24: ...26 ...