DESCRIPTION AND TECHNICAL DATA
6
CAUTIONS
:
•
Do not operate in patient area.
•
Zimmer®
UNIVERSAL compact battery charger (REF 89-8510-421-00) does not meet IEC 60601-1 requirements.
•
Use only with
Zimmer®
UNIVERSAL sterilizable battery (REF 89-8510-440-50) or
Zimmer®
UNIVERSAL battery for Aseptic
Transfer Kit (REF 89-8510-440-20).
•
Check that the voltage of the mains power supply is 100-240 volts.
•
Never block the air vents.
•
Do not immerse compact battery charger.
•
Remove batteries from compact battery charger when recharged (green indicator light).
•
The ON/OFF switch on this unit will not completely shut off all power from the AC outlet. Since the power cord serves as
the main disconnect device for the unit, you will need to unplug it from the AC outlet to shut down all power. Therefore,
make sure the unit has been installed so that the power cord does not create a trip hazard and can be easily unplugged
from the AC outlet if needed.
BATTERY AND ASEPTIC TRANSFER KIT (ATK) HOUSING
REF 89-8510-440-00
89-8510-440-10
89-8510-440-20
89-8510-440-30
Housing
Seal
Lid
ATK Housing
Battery
•
Rechargeable lithium battery
•
Sterilizable housing with detachable
funnel for easy aseptic transfer of
battery into housing
Technical data
Rated Voltage
16.5V DC, 20A
Dimensions
Battery: 76 mm x 60 mm x 40 mm
Housing: 88 mm x 66 mm x 62 mm
Weight
Battery: 230 grams
Housing: 90 grams
Environmental conditions during operation
Ambient temperature
10°C to 38°C (50°F to 100°F)
Relative humidity
30% to 60%
Storage and transportation conditions
WARNING
: Fire and burn hazard. Do not heat above 80°C (176°F) during storage and transportation or incinerate. Exposure
to elevated temperature may reduce battery capacity and performance.
CAUTIONS:
Battery REF 89-8510-440-20
•
Do not sterilize; do not autoclave.
•
Charge only with
Zimmer®
UNIVERSAL Compact Battery Charger (REF 89-8510-421-00).
•
Use only with
Zimmer®
UNIVERSAL Aseptic Transfer Kit Housing (REF 89-8510-440-10).
•
Remove from charger when recharged (green indicator light).
•
Do not immerse or place in automated washer.