DESCRIPTION AND TECHNICAL DATA
7
CAUTIONS:
ATK housing REF 89-8510-440-10
•
Remove ATK housing from handpiece after use.
•
Remove battery from housing before sterilizing.
•
Use only with
Zimmer®
UNIVERSAL battery (REF 89-8510-440-20).
•
Do not wash and sterilize with lid closed.
•
Seal between lid and housing must be present and pressed to guarantee seal.
STERILIZABLE BATTERY
REF 89-8510-440-50
•
Rechargeable lithium battery
•
Sterilizable using FLASH autoclave cycle
Technical data
Rated Voltage
16.5V DC, 20A
Dimensions
88 mm x 66 mm x 62 mm
Weight
310 grams
Environmental conditions during operation
Ambient temperature
10°C to 38°C (50°F to 100°F)
Relative humidity
30% to 60%
Storage and transportation condition
WARNING
: Fire and burn hazard. Do not heat above 80°C (176°F) during storage and transportation or incinerate. Exposure
to elevated temperature may reduce battery capacity and performance. When using heat for sterilization strictly follow the
instructions given in the Cleaning, Decontamination and Sterilization section.
CAUTIONS
:
•
Remove sterilizable battery from handpiece after use.
•
Use only FLASH autoclave cycle (See Sterilization section). Do not exceed 4 minutes at 132°C (270°F) or 3 minutes at 134°C
(273°F).
•
The
Zimmer®
UNIVERSAL Sterilizable Battery is the only component of the
Zimmer®
UNIVERSAL Power System that can be
sterilized using STERRAD® Technology. For details refer to the STERRAD® STERILITY GUIDE.
•
Charge only with
Zimmer®
UNIVERSAL compact battery charger (REF 89-8510-421-00).
•
Remove from charger when recharged (green indicator light).
•
Allow sufficient time for cooling the sterilizable battery after charging in order to introduce it at a normal operating room
temperature in the autoclave.
•
Do not immerse or place in automated washer.
ENGLISH