Oxygen Monitor SGM7.2.4/7.2.6 5 Technical Data
14
5.4
Interface data
Serial interface RS-232
Transfer rate .................................. max. 19200 Baud, adjustable
Stop bits ........................................ 1
Data bits ........................................ 8
Parity ............................................. none
Handshake .................................... without
Pin-configuration connector SUB-D 9-pol. F:
Chart 1:
Serial interface protocol (CR = Carriage Return)
Command
Response
Description
M2CR
M2x.xxExxCR
Oxygen concentration in ppm
A1CR
A1xxxCR
Cell voltage in mV
A2CR
A2xxxCR
Measuring cell temperature in °C
Chart 2:
Error indication
Error code
Description
0
ERROR0
Transmission error
1
ERROR1
Warm-up
2
ERROR2
Cell temperature too low
3
ERROR3
Thermocouple defective
4
ERROR4
Operation hour meter run out!
Change filter or carbon, then reset! (see manual)
5
6
ERROR6
System error
Analog output
SGM7.2.4:
Voltage output, 0/2 - 10V, potential-free, range adjustable
SGM7.2.6:
Current output, 0/4 – 20 mA, potential-free, range adjustable
Chart 3:
Limit value relay
Charge
Resistive charge (cos
ϕ
ϕ
ϕ
ϕ
= 1)
Max. switching voltage
125 V AC, 60 V DC
Max. switching current
1A
Max. switching power
62.5 VA, 30 W
Min. permissible charge
1 mA at 5 V DC
MD
TXD
RXD