Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
24
8.1.5
Status and error messages
During the measuring, the functions of the measuring cell are
monitored. In case of faults and/or errors, messages are displayed.
The relay output is activated simultaneously.
Self-check of the
SGM7.2.4
Chart 6:
Status / error messages
Status
Display
Remark
0
OK
1
LIMIT
2
RANGE <<<
No alarm (only SGM7.2.4)!
Several reasons – see chapter
8.4
3
RANGE >>>
4
FLOW <<<
< 8 l/h
5
FLOW >>>
> 35 l/h
6
Change filter!
> 600 h (or after period adjusted
by customer)
7
8
9
WARM-UP
Cell temperature too low
(wait, after approx. 15 minutes
750 °C are reached)
10
CELL TEMP.<<<
Set temp. -10° and > 30 min
11
ERROR THERMOCOP. Thermocouple broken
12
13
14
SELF-TEST FAILED
SGM7.2.4
Status 1, 4...8:
warnings, only alarm relay active
Status 2, 3:
warnings,
no alarm
by alarm relay
Status 9...14:
failure, alarm relay active
and
analog output
(voltage signal)
10 V
SGM7.2.6
Status 1...8:
warnings, only alarm relay active
Status 9...14:
failure, alarm relay active
and
analog output
(current signal)
0 mA