Oxygen Monitor SGM7.2.4/7.2.6 8 Operation
25
8.2
Parametrization
8.2.1
Adjustable parameters
Chart 7:
Overview of the adjustable parameters
Parameter
Range
Remarks
Display
O
2
: 0…21 vol% or in ppm
Range up to 100 vol% on
request
Analog output
0-10 V or 2-10 V
(SGM7.2.4)
0...20 mA or 4...20 mA
(SGM7.2.6)
Scale of output
Linear
Logarithmic (Basis 10)
Logarithmic scale
recommended if
measuring value exceeds
several decades.
Suppressed zero
point
0.00001…21 vol% or
0.1…210000 ppm
For optimal resolution at a
special range
Max. measuring
value
21.0 vol% or 210000 ppm,
Higher values on request
This value corresponds to
the terminal value of the
analog signal (e.g. 10 V)
Response time t
90
1 ... 99 sec
Formed by arithmetical
average of the measured
value
Limit value
0…99.99 vol%
or 0…999999 ppm
Selectable as lower or
upper limit value with the
signs”>” or ”<”
Delay time of the
limit value
1 ... 99 sec
Time to alarm when limit
value is exceeded
Transmission
capacity of RS-232-
interface
4800, 9600, 19200 Baud
Internal measuring
gas pump
Switch on/off by soft key, switch at
the rear side or RS232
High/Low via keypad or external
controlling (RS232) adjustable
Maintenance interval 0…999 h
Count down, at zero
alarm indication for filter
change