Oxygen Monitor SGM7.2.4/7.2.6 3 Safety regulations
9
3 Safety regulations
The following regulations for industrial safety provide basic information about
potential danger during the operation of the buffer gas monitor SGM7.2.4.
Therefore, they must be observed and strictly followed by the responsible
staff.
•
A failure-free and functional operating of the SGM can only be
guaranteed with knowledge of this manual. Therefore, all chapters of
this manual must be read carefully before the installation and initiation
of the SGM.
•
The SGM is to be used for the functional operation only (see chapter
2).
•
The SGM is to be installed, operated, and serviced by trained staff only.
•
The SGM is to be connected to an isolated ground socket (Schuko-
socket) with the supplied cable.
Explosive gas compounds, halogens in high concentration, and sulphuric
gases (e.g. SO
2
) are not permitted to be measured by this SGM7.2.4.
Due to the high operation temperature of the sensor, the SGM produces a
lot of heat.
Make sure the device does not overheat because of covering.
A vertical installation may cause heat accumulation and
damage to the device.
Switch off and disconnect the device from the power supply before opening
the housing cover of the SGM.
Special safety instructions for potential danger during certain working
processes are given in relevant text passages.