X-Ray Inspection System
User Manual
P a g e
| 63
Copyright©2021 ZKTECO CO., LTD. All rights reserved.
3)
Equipment that needs to be stored for a long time should be packaged and well
stored in a clean, dry and well-ventilated warehouse where there should be no
corrosive gas around and the relative humidity should not be more than 80%.
4)
According to the regulations of measures for the Management of Security
Permission of Radioisotope and Ray Devices issued by the People's Republic of
China, the X-Ray radiation device in this product is Type III ray device. In case of any
issue, the customers are suggested to report to the local competent environmental
department.
5)
When the equipment needs to be discarded, please consult the relevant
professionals or the local competent environmental departments for handling
methods.
Troubleshooting
This section introduces the failures that may occur, causes and troubleshooting methods.
If there are issues that cannot be solved, please contact our professional technician for
consultation and repair.
It is advisable to use the components of our company for maintenance. If any failure or
damage occurs due to unauthorized maintenance or unauthorized components, we will
not be responsible.
Failure 1: The equipment did not boot normally after pressing the Start button.
If the equipment did not boot normally, check whether the power plug is plugged-in
properly or not. Also check if the power supply is normal (200V to 240V). If the voltage
supply is low or abnormal, repair the power supply unit or report to the relevant
department to restore the normal power supply.
1)
If the power supply voltage is normal, but the machine still did not boot, then check
the connection terminal 1 and 3 to see if the voltage between the two terminals is
200V-240V.