Page 29 of 40
RR PARAMETER SETUP
Table 11 on page 38 lists the Default RR Alarm Limits for Adult and Pediatric.
on page 39 lists the Default RR Parameter Setting for Adult and Pediatric. RR Alarm
Limits and Parameter settings will operate for the current monitor patient mode.
RR ALARM LIMITS:
1) Touch RR Numeric field and the Setup RR menu will be displayed (refer to Figure 18);
Figure 18: RR Setup Menu
2) Set Alarms On to Yes to enable RR alarms, or No to disable RR alarms.
Warning:
If Alarms On is set
to “No” the monitor will not generate any visual or audible
indication of an alarm condition for RR.
3) Adjust the desired RR Upper or Lower Limit value.
The RR Upper and Lower Alarm Limits can be adjusted independently.
4) Touch OK to accept or Cancel to ignore the selection.
5) Touch the Home touch area to return to the Main screen.
RR AUTO (set) ALARM LIMITS
The Upper and Lower alarm limits for RR can be set automatically using the Auto feature,
which provides a structure to establish alarms limits based upon your patient’s current
measured value.
To enable Auto limits, open the RR Setup Window and touch the “
AUTO
” key on the same
line with the parameter (refer to Table 3 for RR percentages).
Parameter
Alarm Limit Adjustment
(x% of the current measured value
Lower
Upper
RR
80%
125%
Table 3: RR Auto Alarm Limit Adjustment