VETSCAN SA USER’S MANUAL
v
1.2 SOFTWARE
|
2 2
WARNING!
Ensure the strip has been placed in the tray before the Strip Tray begins to move
into the analyzer.
If a strip is not present in the Strip Tray at the conclusion of the measurement, an error will be
reported on the UA and sent to the SA. The test will be changed to a SA only test.
While measuring, the SA screen is locked, and navigation to other screens is prohibited. Also,
both the SA status indicator light (top right of the SA Results Panel) and the UA status indicator
light (top right of the UA Results Panel) will display yellow during measurement process. Do not
disconnect UA from SA during testing process. If the UA becomes disconnected during testing,
the UA test will be canceled automatically on the SA, and complete as an SA only test. If the
UA becomes disconnected from the SA during the measurement, wait until the SA completes
its measurement to reconnect. If desired, manually send the UA results from the UA to the SA
Database by pushing the “USB send” icon on the UA.
The UA analyzer can operate interdependently from the SA; however, if combined SA and UA
reports are desired, the UA test must be initiated on the SA.
Operating the UA independently from the SA indicates that the test was not initiated from the
SA or that the UA was disconnected from the SA. If UA test results need to be transferred to the
SA, the UA will need to be connected to the SA. Next, navigate to the UA Database and select
the desired test report to transmit and push the “USB send” icon to transmit the data to the SA.