C

HAPTER

1 M

AINTENANCE

 T

ESTS

1–4

www.zoll.com

9650-0309-01 Rev. A

2.0 Preventive Maintenance Test

Use this test to verify that the AED Pro unit is working properly and ready for use.

For an alternative test using the AED Pro simulator, refer to the 

AED Pro Operator’s Guide

.

Note:

To verify that the delivered energy for the first, second, and third shocks corresponds 
to the unit’s preconfigured adult energy levels, use the ZOLL Administration Software 
(ZAS) to obtain the current configuration of the unit.

Tools Needed

Fluke Impulse 4000 Defibrillator/Pacer Performance Analyzer or equivalent

Defibrillation analyzer (universal) adapter cable

Test Setup

Turn off the AED Pro unit and unplug the preconnected electrodes. Keep the electrodes so 
they can be reconnected later.

Connect the universal adapter cable to the AED Pro unit and to the analyzer.

Turn on the analyzer and set its function to VF (ventricular fibrillation).

Variations in Messages

When verifying messages in the test results, please be aware of the following issues:

The wording of some messages is configurable. The following table shows equivalent 
messages.

The message that reports the detected electrode type can be either 

ADULT PADS

PEDIATRIC PADS

, or 

ECG CABLE

.

Some messages are issued only if the unit is configured to do so. If an optional message is 
not issued during a test, do not consider this a failure.

In some cases a text message on the display screen is abbreviated or otherwise differs 
slightly from the associated voice message.

Default Message

Alternative Message

CHECK PATIENT

CHECK RESPONSIVENESS

CHECK PULSE

CHECK CIRCULATION

IF NO PULSE START CPR

IF NO CIRCULATION START CPR

IF NO PULSE CONTINUE CPR

IF NO CIRCULATION CONTINUE CPR

SHOCK ADVISED

TREATMENT ADVISED

NO SHOCK ADVISED

NO TREATMENT ADVISED

PRESS FLASHING SHOCK BUTTON

PRESS FLASHING TREATMENT BUTTON

RELEASE SHOCK BUTTON

RELEASE TREATMENT BUTTON

SHOCK DELIVERED

TREATMENT DELIVERED

NO SHOCK DELIVERED

NO TREATMENT DELIVERED

Summary of Contents for aed pro

Page 1: ...Service Manual...

Page 2: ......

Page 3: ...Service Manual 9650 0309 01 Rev A February 2007...

Page 4: ...o CodeNet stat padz pedi padz pro padz and v pack are registered trademarks of ZOLL Medical Corporation AED Plus PowerCharger Preconnect RescueNet Smart Alarms CPR D padz Rectilinear Biphasic SurePowe...

Page 5: ...tions for Testing 1 2 Test Equipment 1 2 1 0 PhysicalInspection 1 3 2 0 Preventive Maintenance Test 1 4 3 0 BatteryCheck 1 9 4 0 ECG Monitoring Optional 1 9 5 0 ManualSelf Test 1 10 Chapter 2 Troubles...

Page 6: ...TABLE OF CONTENTS ii www zoll com 9650 0309 01 Rev A...

Page 7: ...iew this section thoroughly before servicing an AED Pro unit Chapter 1 Maintenance Tests Provides the recommended procedures for evaluating the condition and performance of an AED Pro unit Chapter 2 T...

Page 8: ...mulated cardiac rhythms Conventions This manual uses uppercase italics for messages that are voiced or displayed on the screen for example DON T TOUCH PATIENT ANALYZING Within text the names and label...

Page 9: ...nts Dangerous voltage Conformit Europ enne Complies with medical device directive 93 42 EEC Keep away from open flame and high heat Do not open disassemble or intentionally damage Do not crush Nonrech...

Page 10: ...tions Only qualified personnel should service or disassemble an AED Pro unit Before using servicing or disassembling this device read the AED Pro Operator s Guide Before servicing or disassembling any...

Page 11: ...s service contact the ZOLL Technical Service Department Telephone 1 800 348 9011 U S A customers only 1 978 421 9655 Fax 1 978 421 0010 When requesting service please provide the following information...

Page 12: ...cal Service Department SR number Telephone 1 866 442 1011 In other locations The nearest authorized ZOLL Medical Corporation representative To locate an authorized service center contact the Internati...

Page 13: ...of which requires a pass fail rating For each step perform the specified action and then verify that the expected results occur If all expected results occur assign a Pass rating to the step otherwise...

Page 14: ...re optional ZOLL AED Pro Simulator When connected to an AED Pro unit the simulator can send it various simulated cardiac rhythms ventricular fibrillation normal sinus rhythm or asystole and optional s...

Page 15: ...of wear reasonable and not excessive 1 3 Are the front panel and housing undamaged and free of cracks 1 4 Are input connectors clean and undamaged 1 5 Are electrodes preconnected is the cable free fr...

Page 16: ...ion to VF ventricular fibrillation Variations in Messages When verifying messages in the test results please be aware of the following issues The wording of some messages is configurable The following...

Page 17: ...TOUCH PATIENT ANALYZING TREATMENT ADVISED or SHOCK ADVISED DON T TOUCH PATIENT PRESS FLASHING TREATMENT BUTTON or PRESS FLASHING SHOCK BUTTON 2 2 Press the Shock button first shock Analyzer indicates...

Page 18: ...ED or SHOCK ADVISED DON T TOUCH PATIENT PRESS FLASHING TREATMENT BUTTON or PRESS FLASHING SHOCK BUTTON 2 4 Press the Shock button third shock Note If your AED Pro unit is not configured for stack shoc...

Page 19: ...G GIVE TWO BREATHS CHECK CIRCULATION or CHECK PULSE IF NO CIRCULATION START CPR or IF NO PULSE START CPR 2 6 Lift the puck that is attached to the universal adapter cable Move it up and down very slig...

Page 20: ...nit Disconnect the universal adapter cable from the AED Pro unit 2 8 Wait 10 seconds and then turn on the AED Pro unit PLUG IN CABLE CHECK RESPONSIVENESS or CHECK PATIENT CALL FOR HELP These three pro...

Page 21: ...None Test Procedure Test Action Expected Result s Pass Fail N A 3 1 Open battery door and verify that the battery has no signs of visible damage 3 2 For a sealed lead acid battery verify that the bat...

Page 22: ...tware are functional CPR monitoring circuitry and compression depth sensor are functional when ZOLL CPR D padz are attached Audio output circuitry is functional Display screen is functional Tools Need...

Page 23: ...steps to solve the problem This chapter also lists AED Pro error log messages that indicate the unit requires technical service This chapter contains the following Troubleshooting the AED Pro Unit on...

Page 24: ...the unit continues to fail take it out of service Contact ZOLL Technical Service Unexpected shutdown in clinical mode Note In clinical mode the unit automatically powers off if it does not detect a pa...

Page 25: ...S FLASHING SHOCK BUTTON before pressing the button None PLUG IN DEFIB CABLE Note This prompt appears if an ECG cable is connected but the unit is not configured for ECG monitoring Check the cable and...

Page 26: ...ine Turn off nearby two way radios and cell phones Before attaching electrodes properly prepare the patient s skin refer to the AED Pro Operator s Guide Ensure proper adhesion of the electrodes to the...

Page 27: ...or again and deliver the shock while the charge ready tone sounds The Shock button was pressed before the unit was fully charged Wait for the charge ready tone and a flashing Shock button before press...

Page 28: ...MUNICATIONS ESTABLISHED refer to the ZAS help for instructions on saving the unit s history file and viewing the included error log Table 2 4 AED Pro Error Log Messages Error No Error Message 1 The pa...

Page 29: ...curred when calling a Nucleus operation 21 An unexpected error has occurred when calling a Rhapsody operation 22 An unexpected coding error has occurred 23 No Language installed corrupt language 24 Th...

Page 30: ...CHAPTER 2 TROUBLESHOOTING 2 8 www zoll com 9650 0309 01 Rev A...

Page 31: ...appendix provides a sample maintenance checklist for the AED Pro unit To ensure that the equipment is functioning properly and ready for emergency use complete the checklist periodically according to...

Page 32: ...structions To maintain records for maintenance testing photocopy the checklist shown on the following pages The checklist includes an entry for each test described in this manual Use these copies to r...

Page 33: ...and keep it for your records For a description of each numbered test refer to Chapter 1 Maintenance Tests in the AED Pro Service Manual Serial No Location Date Tester Signature Result of Check No acti...

Page 34: ...A 4 www zoll com 9650 0309 01 Rev A...

Page 35: ...rompts 1 4 Conventions used in manual iv D Defibrillator analyzer 1 2 1 4 1 9 Defibrillator troubleshooting 2 5 Documentation iv E ECG troubleshooting 2 4 Error log 2 6 H History file device 2 6 I Imp...

Page 36: ...erations vi Self test manual 1 10 Service technical vii Simulator iv 1 2 Software ZOLL Administration 1 2 2 6 Symbols on equipment v T Technical Service vii Tests maintenance 1 1 checklist A 1 prerequ...

Reviews: