AED Pro Error Log Messages
9650-0309-01 Rev. A
ZOLL AED Pro Service Manual
2–7
14
Patient impedance experienced a short
15
The A2D is getting non-accurate readings during its test
17
The message queue overflowed
18
PowerOn task has timed out while waiting for a self test result to come back.
19
Calibration faults that occur during a clinical event.
20
An unexpected error has occurred when calling a Nucleus operation.
21
An unexpected error has occurred when calling a Rhapsody operation.
22
An unexpected coding error has occurred.
23
No Language installed - corrupt language.
24
The shock button is pressed/stuck.
25
The ECG Calibration data is invalid
26
The Patient Impedance Calibration data is invalid
27
A critical flash error has occurred
28
The WatchDog is not enabled during clinical event
29
The clock test failed
Note:
Error numbers 256 and greater indicate noncritical conditions.
Table 2-4. AED Pro Error Log Messages (continued)