906-0731-01-05 Rev. C
ZOLL Ventilator Operator’s Guide
A-3
Pulse Oximeter
Range
Saturation (% SpO
2
)
1%-100%
Pulse Rate (bpm)
25-240
Perfusion
0.02%-20%
Accuracy
Saturation (% SpO
2
)-During No Motion Conditions
Adults, Pediatrics
70%-100%
2 digits
0%-69% unspecified
Neonates
70%-100%
3 digits
0%-69% unspecified
Saturation (% SpO
2
)-During Motion Conditions
Adults, Pediatrics
70%-100%
3 digits
0%-69% unspecified
Neonates
70%-100%
3 digits
0%-69% unspecified
Pulse Rate (bpm)-During No Motion Conditions
Adults, Pediatric, Neonates 25 to 240
3 digits
Pulse Rate (bpm)-During Motion Conditions
Adults, Pediatric, Neonates 25 to 240
5 digits
Resolution
Saturation (% SpO
2
) 1%
Pulse Rate (bpm) 1
Low Perfusion Performance
>0.02% Pulse Amplitude Saturation (% SpO
2
)
2 digits
and% Transmission >5% Pulse Rate
3 digits
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal
to the amount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that change
usual arterial pigmentation may cause erroneous readings.