S
PECIFICATIONS
A-4
www.zoll.com
906-0731-01-05 Rev. C
Device Classification
The following table describes the ZOLL Ventilator’s device classification:
Category
Classification
Type of Protection against Electric
Shock
The medical power supply (which contains the
system’s safety barrier) is labeled as Class I or Class
II. Electrical shock protection is not dependent upon
earthing since this power supply design includes
double insulation.
Degree of Protection against
Electric Shock
Applied Parts
The ventilator circuit is Type BF applied part.
The pulse oximeter is Type BF Defibrillation Proof
Applied Part.
Degree of Protection against
Harmful Ingress of Water
IPX4: Splash-proof equipment rating, include:
•
Padded case with rain flap
•
Bacterial/viral filter to protect the compressor
Method of Sterilization or
Disinfection
O2 Supply hoses and connections should be wiped
with a damp, soapy cloth and thoroughly dried with a
lint-free cloth. The unit’s housing should also be
cleaned as necessary with a damp, soapy cloth and
throughly dried with a lint-free cloth. Do not clean
with abrasives or chlorinated hydrocarbon cleansers.
Ventilator circuits are only for single use. Follow all
IFU instructions.
Degree of Safety of Application in
the presence of a Flammable
Anesthetic Mixture with Air or with
Oxygen or Nitrous Oxide
Equipment
not
suitable for use in presence of
Flammable Anesthetic Mixture of Air or with Oxygen
or Nitrous Oxide
Mode of Operation
Continuous Operation