906-0731-01-05 Rev. C
ZOLL Ventilator Operator’s Guide
2-7
Display Screen
The ZOLL Ventilator’s display screen has four functional areas:
Message Area
The display screen’s message area can display the following:
•
Airway Pressure and Pleth Waveform Plots
-- Under normal operation (as in the example
above), the message area displays plots for airway pressure and, when the pulse oximeter is
connected, the Pleth waveform.
•
Menus
-- Displays the Main Menu after you press the menu button on the ventilator’s
control panel, or displays a parameter’s context menu (which appears after you
press and
hold
the associated parameter button on the control panel). When a plot is necessary to
facilitate a parameter adjustment, the message area displays both the plot and the
parameter’s context menu.
•
Alarms
-- When an alarms occur, the message area displays Smart Help messages that
identify the alarms and describe possible causes and actions that you can take in response.
•
Popup Windows
-- Display information that assists you when adjusting parameter values.
Parameter Windows
Parameter windows display the measurements, alarm limits, and associated parameters for their
labeled parameters. Parameter values that you can adjust, such as alarm limits, appear as solid
text. Parameter values that you cannot adjust, such as measurements taken by the ventilator,
appear as outlined text. We provide information on adjusting parameter values in Chapter 4,
“Using the ZOLL Ventilator.”
Message Area
Shared Icon Area
Auxiliary Parameter Boxes
Parameter Windows