Fresh Gas Intake and Attachments
The fresh gas intake, which is located on the side of the ventilator, allows ambient air into the
unit’s internal compressor. The intake also acts as an anti-asphyxia value that enables the
patient to breath ambient air should the ventilator fail.
The fresh gas intake contains a particulate filter and permits the operator to connect either a
bacteria/viral or a chemical/biological filter depending on ambient conditions
ZOLL provides an O2 Reservoir Kit to allow for low flow oxygen supply to the ventilator. An
oxygen concentrator source provides oxygen to the O2 reservoir.
Top Panel
The ZOLL Ventilator’s top panel appears as follows:
The oxygen hose, ventilator circuit, external power, and pulse oximeter attach to the top panel
of the ventilator. The USB port is only used when servicing the unit.
External Power Input
Gas Output
High Pressure
Oxygen Input
Transducer
(Patient Airway
Pressure)
Exhalation
Valve
Fresh Gas/
Emergency
Air Intake
Pulse Oximeter Connector
+ USB Connector