ZOLL Ventilator Operator’s Guide
3-3
2. Attach the High Pressure Oxygen Supply (Optional)
Since the ventilator includes an internal compressor, the attachment of a high pressure oxygen
supply is optional.
Review the high pressure supply requirements that we describe in Chapter 2, and use the
oxygen hose to attach the ventilator’s oxygen inlet to the high pressure gas source.
Note:
Use only with medical-grade (USP) oxygen. When using with an oxygen cylinder, the
cylinder must be secured.
Note:
The O
2
Hose is either colored green or white, depending on country specifications.
3. Inspect Fresh Gas Intake Filters
The fresh gas intake is the gas source for the ventilator’s internal compressor. The ventilator
normally operates with two built-in filters:
1. Removable Foam Filter (
REF
465-0028-00)
2.
Fresh Gas Intake Disk Filter (
REF
465-0027-00)
Inspect the filters and, if dirty, replace them (See Chapter 7, “Replacing the Ventilator’s
Filters”).
Oxygen Inlet
Fresh Gas Intake