Breath Target
The selected ventilation mode, and the selection of breath target (volume or pressure)
predetermines the parameter availability for the BPM, Vt, and PIP parameter windows.
Volume targeting
assures a constant volume is delivered to the patient in the inspiratory time
using a constant flow.
Pressure targeting
provides a constant airway pressure for the duration of the inspiratory time
Leak Compensation
Leak Compensation provides flow during the expiratory phase to maintain the baseline
pressure in patients that are breathing spontaneously, but have a leaking airway or facemask
To avoid nuisance alarms in patients with active leaks, Leak Compensation suppresses he
following alarms:
•
Incomplete Exhalation (Alarm #3091)
•
Insufficient Flow (Alarm #2095)
The following table lists the ventilator modes and their availability in the ZOLL ventilator
models, and gives the options and ranges for the ventilation mode parameters:
Parameter Window
Options /
Range
Availability/Notes
Models
Primary Value
Mode
AC
All
SIMV
EMV+, Eagle II
CPAP
All
BL
All
Secondary
parameter Val-
ues
Target
(V) or (P)
AC and SIMV modes
All
(SIMV not available
with AEV)
LC
On or OFF
Default off
Default on
AC(P), SIMV(P) modes
EMV+, Eagle II
BL modes
All
CPAP
Alarms
N/A
Measured Value N/A
Apnea Back Up
Context Menu
CPAP and BL modes
Apnea Back Up
BPM
1 to 80
All
Apnea Back Up
PIP
10 to 80
All
Apnea Back Up
I:E , Ti
1:1 to
1:99, 0.1 to 3
Control selected in con-
text BPM Context Menu
All