U
SING
THE
ZOLL V
ENTILATOR
4-16
www.zoll.com
906-0731-01-05 Rev. C
HR (Heart Rate)
The HR (Heart Rate) parameter window displays the patient’s heart rate when the pulse
oximeter is working and the sensor is attached.
Pressing the
HR
parameter button highlights the High Heart Rate alarm limit and enables its
value to be changed. Pressing the
HR
button a second time highlights the current value of the
Low Heart Rate Alarm limit and enables its value to be changed. Both limits are adjustable by
1 b/min. The default value at start up for the high alarm limit is 120 BPM (Beats Per Minute);
the low alarm limit is 40 BPM.
The following table gives the options and ranges for the HR parameter:
Parameter Window
Options /
Range
Availability/Notes
Models
Primary Value
HR
%
0 to 240
Measurement - Heart Icon blinks
at the beat rate.
All
Secondary
Values
Not Applicable
Alarm Limits
High Limit
80 to 240,
Off
Low Limit
30 to 79,
Off
Measured
Values
Pleth Waveform
HR
Context Menu
(note same as SpO2 Context Menu)