Page 5-2
User Guide
Figure 5-2
Replace Battery Screen
5.2
Troubleshooting User Advisories and Faults
The Platform enters a User Advisory state (see Figure 5-3) or the Fault State (see Figure 5-4) when one
of several conditions is detected. A User Advisory generally indicates that a misalignment or
inappropriate movement of the patient or the LifeBand has occurred. A Fault generally indicates that the
Platform has detected an inappropriate internal condition. Both conditions are typically correctable by
the operator. Follow the instructions on the screen and then attempt to RESTART active operation by
pressing the Start/Continue button. If that does not work you should follow these general steps to
troubleshoot Advisories and Faults:
1.
Check for correct patient alignment (refer to Section 3.1, “Deploying the AutoPulse System” for
more information), fully extend LifeBand and attempt to RESTART active operation by pressing
the Start/Continue button.
2.
If the user advisory or fault persists:
a)
Remove the LifeBand from, and fully re-insert into, the Platform (refer to Section 2.1,
“LifeBand Load-distributing Band”) and then press the Start/Continue button again.
b) Remove and replace the LifeBand with a new LifeBand and then press the Start/Continue
button again.
c)
Check the Platform for blocked vents.
3.
If a User Advisory or Fault indicator cannot be cleared, record the User Advisory or Fault number
and contact ZOLL at +1.800.348.9011 or +1.978.421.9655.
R E P L A C E B A T T E R Y
m e n u
R e p l a c e / r e c h a r g e b a t t e r y .
R E S TA R T
C O N T I N U O U S
Summary of Contents for AutoPulse 100
Page 1: ...P N 12555 001 Rev 9 Resuscitation System Model 100 User Guide...
Page 8: ...Page viii User Guide This page left intentionally blank...
Page 38: ...Page 2 16 User Guide This page left intentionally blank...
Page 66: ...Page 4 6 User Guide This page left intentionally blank...
Page 72: ...Page 5 6 User Guide This page left intentionally blank...
Page 74: ...Page A 2 User Guide This page left intentionally blank...