2SHUDWLQJ&RQWUROVDQG,QGLFDWRUV

The ZOLL Base PowerCharger

4x4

 with QuickCharge 

illuminates the 

Batt. Ready

 indicator when the battery is 

fully charged. The full charging cycle is complete in 
four (4) hours or less for the PD 4410/Smart/Smart 
Ready Battery and eight (8) hours or less for all XL 
Batteries. The QuickCharge device provides no 
information regarding battery capacity because it does 

not 

test the battery capacity with each recharge. The 

ZOLL Base PowerCharger

4x4

 with QuickCharge relies 

on the user to manually initiate the battery test cycle at 
recommended intervals (see “Manual Battery Testing 
Procedure” on page 9)
 in order to confirm adequate 
battery capacity.

 32:(5

Illuminates when the ZOLL Base PowerCharger

4x4

 is 

connected to live AC mains and the unit is ready for 
charging and/or testing. 

 &+$5*(521

Illuminates when a battery is properly installed in the 
corresponding charging/testing compartment and the 
charge cycle is in progress. 
There are four 

CHARGER ON

 indicator lights. Each one 

corresponds to the battery charging/testing compartment 
directly above it. 

 %$775($'<4XLFN&KDUJH

Illuminates at the end of a charge cycle. Indicates that 
the battery has been charged to 100% of its present 
capacity and is ready for use. The 

BATT. READY 

light 

does NOT give any indication as to the available run time 
left in the battery.
Illuminates at the end of a battery test or battery 
compartment test and remains lit as long as battery 
remains in compartment. Indicates the battery 
compartment or battery has passed the capacity test and 
is operating properly. 
There are four 

BATT. READY

 indicator lights. Each one 

corresponds to the battery charging/testing compartment 
directly above.

 )$8/7

Illuminates, with the 

TESTING

 light, when a battery 

capacity test fails. Also, illuminates when a battery 
compartment or any other failure occurs.
There are four 

FAULT

 indicator lights. Each one 

corresponds to the battery charging/testing compartment 
directly above.

 7(67,1*

Illuminates when either a battery capacity test cycle is in 
progress or a battery compartment test is completed. 

Illuminates with the 

BATT. READY

 light indicating a 

usable fully charged battery, capable of powering an 
M Series equipped with Pace/Defibrillation and Pulse 
Oximetry (SpO

2

) for approximately 1.5 hours in Monitor 

mode.
Illuminates with the 

FAULT 

light to indicate that a battery 

has failed the capacity test. 
There are four 

TESTING

 indicator lights. Each one 

corresponds to the battery charging/testing compartment 
directly above.

 7(67EXWWRQ

Initiates a battery test cycle or battery compartment test. 
The battery capacity test cycle automatically charges, 
discharges and recharges the battery. 
If a battery is installed in the battery compartment and 
the corresponding 

TEST

 button is pressed, a battery 

capacity test is performed. If the battery compartment is 
empty and the corresponding 

TEST

 button is pressed, a 

compartment test is performed and an audible tone will 
sound, indicating that the test has passed.
There are four 

TEST 

buttons. Each one corresponds to 

the battery charging/testing compartment directly above.

6HWXS

Place the ZOLL Base PowerCharger

4x4

 on a stable 

secure surface. There must be at least four (4)  inches of 
clear space around the unit to allow for proper heat 
dissipation during charging or testing batteries. Do not 
place anything on top of or beneath the unit, such as 
blankets or cloths. Doing so may block the vents on the 
unit, preventing proper dissipation of heat during 
operation.
Turn the unit on by plugging the line cord into the power 
module at the rear of the unit and then into live AC 
mains. 

%DWWHU\&KDUJLQJ3URFHGXUH

4XLFN&KDUJH

The ZOLL PD 4410/Smart/Smart Ready Battery may be 
charged in any of the four battery compartments. 
XL Batteries should only be charged in a ZOLL Base 
PowerCharger

4x4

 labeled “XL Battery Ready”. Each 

battery charge compartment operates independently of 
the others. Four batteries may be charged 
simultaneously. Installation or removal of adjacent 
batteries will not affect the ongoing charging of batteries 
in other compartments. Correct placement of the battery 
into any compartment will automatically initiate the 
charging cycle.

Summary of Contents for Base Powercharger 4x4

Page 1: ...ZOLL Base PowerCharger4x4 Operator s Manual Base PowerCharger 4x4 TM CAUTIO N DO NOT BLOCK VENTS OR PLACE ANYTHING ON TOP OF THE VENTS HEAT DAMAGE MAY OCCUR Part Number 9650 0054 Rev G...

Page 2: ...North America Consult your purchasing agreement for terms and conditions associated with your warranty Outside of North America consult the ZOLL authorized representative In order to maintain this wa...

Page 3: ...UH DWWHU RPSDUWPHQW 7HVWLQJ 3URFHGXUH DXOW RGHV DQG HILQLWLRQV 4XLFN 5HIHUHQFH KDUW IRU 2 DVH 3RZHU KDUJHU 4x4 ZLWK 4XLFN KDUJH 6HFWLRQ 2SHUDWLRQ XWR7HVW 2SHUDWLQJ RQWUROV DQG QGLFDWRUV 6HWXS DWWHU KD...

Page 4: ...ZOLL Base PowerCharger 4x4 Operator s Manual ii This page intentionally left blank...

Page 5: ...er 4x4 with AutoTest automatically tests battery capacity with each battery recharge In addition the ZOLL Base PowerCharger 4x4 with AutoTest illuminates the Batt Ready indicator when the battery is f...

Page 6: ...HVKRRWLQJ This section describes several routine checks to assure proper unit operation It includes cleaning and maintenance instructions and troubleshooting guides intended for use by non technical p...

Page 7: ...This warranty does not cover items subject to normal wear and burnout during use including but not limited to lamps fuses batteries patient cables and accessories d The foregoing warranty constitutes...

Page 8: ...epeated use of partially charged batteries will reduce their capacity and operating life Do not use a battery unless the BATT READY indicator is lit and the FAULT light is off Failure to do so may res...

Page 9: ...compartment FAULT Illuminates when battery capacity test or battery compartment test fails Separate indicator light for each battery compartment TESTING Illuminates when either a battery test or a bat...

Page 10: ...order to allow tracking of loan equipment You will be given a Return Claim Number under which to return the ZOLL Base PowerCharger 4x4 Send the unit in its original packaging to ZOLL Medical Corporati...

Page 11: ...x4 refer to QuickCharge unless otherwise specified If your device does not indicate QuickCharge or AutoTest on the label in the upper right corner refer to the QuickCharge section of this manual for p...

Page 12: ...1 Illuminates when either a battery capacity test cycle is in progress or a battery compartment test is completed Illuminates with the BATT READY light indicating a usable fully charged battery capab...

Page 13: ...nsecutive beeps to indicate that the battery is not fully charged and should not be used In addition all four indicator lights will illuminate for four 4 seconds and then go off Repeated use of partia...

Page 14: ...tor lights will illuminate for four 4 seconds and then go off DWWHU RPSDUWPHQW 7HVWLQJ 3URFHGXUH The ZOLL Base PowerCharger 4x4 has four battery compartments that are able to test and charge batteries...

Page 15: ...d range Low Constant Current Amount of current during a charge was below the specified range High Constant Current Amount of current during a charge was above the specified range VCC Voltage Internal...

Page 16: ...e ZOLL Base PowerCharger 4x4 will beep to indicate the battery has been removed and should NOT be used The POWER light should be on as long as the unit is plugged into a live AC source 23 5 7 21 4XLFN...

Page 17: ...r 4x4 refer to AutoTest unless otherwise specified If your device does not indicate QuickCharge or AutoTest on the label in the upper right corner refer to the QuickCharge section of this manual for p...

Page 18: ...tment or any other failure occurs There are four FAULT indicator lights Each one corresponds to the battery charging testing compartment directly above 7 67 1 Illuminates when either a battery capacit...

Page 19: ...cific defect see Fault Codes and Definitions on page 17 and take corrective action Before disposing of the battery test the battery compartment for defective operation see Battery Compartment Testing...

Page 20: ...ry will cause an audible warning of five 5 consecutive beeps to indicate that the battery is not fully charged and should not be used All four indicator lights will illuminate for four 4 seconds and t...

Page 21: ...range Low Constant Current Amount of current during a charge was below the specified range High Constant Current Amount of current during a charge was above the specified range VCC Voltage Internal fi...

Page 22: ...ger 4x4 will beep to indicate the battery has been removed and should NOT be used The POWER light should be on as long as the unit is plugged into a live AC source 23 5 7 21 XWR7HVW 5 5 21 77 5 8 7 7...

Page 23: ...all battery compartments empty check that all indicator lights except POWER are off 3 Press the TEST button under each compartment to verify that the compartments are fully operational see Battery Com...

Page 24: ...r power source Check for blown fuse 2 CHARGER ON indicator light does not illuminate when battery is in compartment Check that battery is properly seated in compartment Check that battery contacts are...

Page 25: ...ent refer to the ZOLL Battery Management Program part number 9650 0019 2 DWWHU 3DFNV ZOLL PD 4410 Smart Smart Ready Battery and XL Battery Packs are five cell assemblies of sealed lead acid batteries...

Page 26: ...ZOLL Base PowerCharger 4x4 Operator s Manual 22 This page intentionally left blank...