Alarms & Corrective Actions
4
Alarm Message
Cause
Corrective Action
DOWNLOAD ERROR.
PLEASE CHECK
EXTERNAL
COMPUTER
Patient trend data download
has failed or there is no
computer connected to the
System.
Check the connection
between the computer and the
System and select “Try again.”
If problem persists, select
“Cancel” and contact ZOLL
technical support.
PATIENT DATA LOG
FULL. PLEASE
DOWNLOAD NOW
Patient trend data log is full.
Only two hours of space
remain.
Download patient trend data.
If data is not downloaded, it
will be overwritten.
Hi PATIENT
TEMPERATURE
ALARM
The patient’s temperature is
above the programmed
Hi
patient temperature alarm
value.
The System alarm will reset
when the patient’s
temperature falls below the
programmed
Hi
patient
temperature alarm value.
Lo PATIENT
TEMPERATURE
ALARM
The patient’s temperature is
below the programmed
Lo
patient temperature alarm
value.
The System alarm will reset
when the patient’s
temperature rises above the
programmed
Lo
patient
temperature alarm value.
PRIMARY
TEMPERATURE
PROBE
DISCONNECTED*
The primary patient
temperature probe is
disconnected.
Verify position of primary
temperature probe. Assure
that primary probe is plugged
into socket T1, or if an
interconnect cable is used, the
probe is plugged into the cable
and the cable is plugged into
T1.
PRIMARY
TEMPERATURE
PROBE DISLODGED*
Temperature output from the
primary patient temperature
probe changed by –0.2° C or
more within 10 seconds.
Replace probe in patient.
PUMP DOOR OPEN*
The pump cover is open.
Close the cover.
PUMP FAILURE*
The pump has failed.
Inspect the pump. Clear any
obvious faults and restart. If
the problem persists,
discontinue use and contact
your ZOLL service
representative (see
Chapter
9
).
600479-001 Rev 2 123/147
Summary of Contents for CoolGard 3000
Page 1: ...Operation Manual 600479 001 Rev 2 1 147...
Page 4: ...Safety Information 1 Safety Information 600479 001 Rev 2 4 147...
Page 13: ...Introduction 1 Introduction 600479 001 Rev 2 13 147...
Page 25: ...Receiving Inspection Assembly 1 Receiving Inspection Assembly 600479 001 Rev 2 25 147...
Page 32: ...Operation 1 Operation 600479 001 Rev 2 32 147...
Page 95: ...Accesories HMIA 1 Hospital Monitor Interface Accessory HMIA 600479 001 Rev 2 95 147...
Page 109: ...TempTrend CSV 1 TempTrend CSV 600479 001 Rev 2 109 147...
Page 120: ...Alarms Corrective Actions 1 Alarms Corrective Actions 600479 001 Rev 2 120 147...
Page 125: ...Troubleshooting 1 Troubleshooting 600479 001 Rev 2 125 147...
Page 131: ...Maintenance 1 Maintenance 600479 001 Rev 2 131 147...
Page 140: ...Warranty Service 1 Warranty Service 600479 001 Rev 2 140 147...
Page 145: ...Specifications 1 Specifications 600479 001 Rev 2 145 147...