Operation
39
Figure 4-45. Making a Sterile Connection to the Saline Container.
CAUTION! START-UP KIT SPIKE
The spike on the Start-up Kit is relatively long. Be careful not to puncture the
side wall of the saline bag when connecting to the Start-up Kit.
Prime & Fill the Air Trap
16.
Remove the air trap from its holder and hold it upside down (with the
tubing connections pointing downward).
Figure 4-46. Hold the Air Trap Upside Down.
17.
Prime the air trap and the tubing circuit by pressing and holding the
PRIME
switch. The roller pump will slowly start and take about 20