58
2. If the primary and secondary patient temperature probes have not been placed in the patient, this should
be done now. Refer to the Instructions for Use that accompany the temperature probes for information
about the probes.
3. Connect the blue patient temperature cable to the YSI-400 primary temperature probe (e.g. Foley
catheter, rectal or esophageal). Connect the plug at the end of the blue patient temperature cable into
the connector labeled “T1” on the front of the console.
Figure 4.48. Temperature Probe Connections
4. If you are using a secondary patient temperature probe, connect the blue patient temperature cable to the
YSI-400 secondary temperature probe. Connect the plug at the end of the blue patient temperature cable
into the connector labeled “T2” on the front of the console. If you are not using a secondary temperature
probe, the patient MUST be monitored by a separate hospital patient temperature monitor.
5. Place the ZOLL catheter in the patient now. Refer to the Instructions for Use for information about the
catheter.
6. The tubing to the catheter is supplied with the supply and return connectors connected to each other.
Using aseptic technique, disconnect the two connectors.
Figure 4.49. Disconnect the Connectors Using Aseptic Technique
7. Connect the female tubing connector to the male connector on the catheter. The gender of the tubing
connectors and the catheter connectors ensures that they cannot be connected backwards.