Responding to alarms
Page 5-21
Too much gel
After you add gel to the electrode belt, you may see this message telling
you there is too much gel between the therapy electrodes.
Follow the instructions below to remove the excess gel.
While you’re wearing the LifeVest, tap the help button
for reminders
about what to do.
Tap
OK
to resume normal monitoring.
To remove excess gel
1 Remove the battery from the monitor.
2 With the electrode belt and garment on your body, and using a towel or soft
cloth, wipe your skin between the front and back therapy pads on the left side
of your body. Be careful not to remove the gel under the therapy pads.
3 Reinstall the battery into the monitor.