Date:  February 8, 2021 

By:  Zoll  Chelmsford, MA  USA 
File =  Zoll Guideline 

DISCLAIMER:  THIS PROCEDURE PROVIDED "AS IS" AND WITH 
POSSIBLE FAULTS.   USER MUST VERIFY BEFORE USE.  NEITHER 
PROVIDER NOR WEBSITE ASSUMES ANY RESPONSIBILITY FOR ITS USE. 

 
1.

 

General 

Service Guideline for Zoll Series M and E with ECG, SP02, NIBP. 

 
2.

 

Reference Documents 

Operator Manual M Series 9650-0200-01 
Service Manual M Series  9650-0450-01 
Configuration Guide M Series  9650-0201-01 
 
Operator Manual E  9650-1210-06 
Service Manual E Series 9650-1450-01 
Configuration Guide E Series 9650-1201-01 

 

3.

 

Accessories (see Appendix for photos) 

“MFC” HV Paddle Cable, Hard Paddles or “Padz® Adult” See Appendix

 

Recorder ECG Paper

 

(90 x 90 x 200 Z-fold)

  

ECG Cable: 3 lead, 5 lead (cable uses normal 6-pin connector) 
SP0

2

 Sensor/Cable (Massimo normally supplied) 

CO

2

 Sensor/Cable, and data card – not addressed here. 

NIBP Cuff/Tubing – standard issue from various sources. 
 

4.

 

Test Fixtures 

ECG Simulator; Defibrillator Analyzer if the Dummy Load not available. 

 
Note:  Some units are manual mode only (no AED option); other units have both modes and can be configured to turn-on 
either in manual mode, or AED mode.  

The following checks are written with the unit in manual mode, which is the 

factory default.  Other various user options are beyond the scope of this document.  Warning: Performing a 
“factory defaults”(a reset) to select manual mode at turn-on, will affect any custom configurations.

 

 

5.

 

Basic PM Procedure  (every 6 months per Zoll)  
 
5.1.

 

Physical Inspection

 – 

Note paddles, case & power cord integrity, and battery age.  Does handle works 

properly?  Recorder drawer opens and closes properly?  Input connectors clean and undamaged?  No 
cuts on the accessory cables.

 

 

5.2.

 

Turn On / Self Test 
After connection to AC power, note the CHARGER ON indicator. Turn the defibrillator ‘ON’ by turning 
the rotary knob on the main panel. 

 

 

5.3.

 

Battery  Diagnostics / Info:   
The E Series products use easily replaced sealed lead-acid or lithium-ion battery packs when new and 
fully charged.  They provide at least 2.5 hours of monitoring.  M-Series use only sealed lead-acid. 
The amber or green lights (below the screen) will illuminate after plugging the instrument to the mains. 
If both lights flash ON/OFF, no battery is installed or the unit is defective.  With a new battery, the green 
LED usually lights within a few minutes.   

 

Note:  For a Zoll M or E Series compatible battery (10V/2.5Ah), one verified item is R&D Batteries, item #5603 – 
sealed lead acid. 

 

 

Zoll M 

 

Zoll E 