Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)
7-8
www.zoll.com
9650-0806-01 Rev. D
Extreme Bradycardia and Extreme Tachycardia Alarm Limits
The Extreme Brady/Tachy Alarm Settings menu specifies the configured delta values that are
applied to the HR lower limit alarm and HR upper limit alarm for the Extreme Bradycardia and
Extreme Tachycardia LTA alarm detection, respectively. For example, if the HR lower limit
alarm is configured to 50 and the Extreme Brady delta value is configured to -10, the Extreme
Bradycardia LTA detection threshold will be at 40 bpm. Likewise, if the HR upper limit alarm
is configured to 120 and the Extreme Tachy delta value is configured to +20, the Extreme
Tachycardia LTA detection threshold will be at 140 bpm. The HR lower and upper limit alarms
are not required to be enabled for the Extreme Bradycardia and Extreme Tachycardia LTA
alarm detection to be enabled. The Extreme Bradycardia and Extreme Tachycardia LTA alarms
will not activate until 20 seconds after the alarm limit threshold is crossed.
The following tables list the default Extreme Bradycardia and Extreme Tachycardia alarm
limits for adult, pediatric, and neonate patients, and the ranges in which you can set these
limits:
Extreme Bradycardia Alarm Limits
Extreme Tachycardia Alarm Limits
Acknowledging LTA Alarms
LTA alarms are latching, which means that the alarm indication will persist until both of the
following conditions are met:
1. The alarm has been acknowledged by pressing the Alarm Silence/Reset button on the front
panel of the unit.
2.
The alarming condition no longer exists.
Patient Type
Default
Delta Value Range
Adult
-10
-5 to -30 in increments of 5
Pediatric
-10
-5 to -30 in increments of 5
Neonate
-10
-5 to -30 in increments of 5
Patient Type
Default
Delta Value Range
Adult
+20
+5 to +30 in increments of 5
Pediatric
+20
+5 to +30 in increments of 5
Neonate
+20
+5 to +30 in increments of 5
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...