Do not use NIBP without proper training. Check that the correct patient mode has
been selected to ensure that the initial inflation pressure is set correctly. Patient
movement, very low pulse volume, or vibration from outside sources can influence
the accuracy of blood pressure measurements.
•
Do not attempt to take NIBP measurements on patients during cardiopulmonary
bypass procedures.
•
Some or all NIBP safety functions are disabled when performing the NIBP test in the
Service menu. Do not conduct NIBP tests when the cuff is attached to a patient.
How does NIBP Work?
The blood pressure cuff and hose connect to the Propaq MD unit through the NIBP connector
on the side panel of the unit. The NIBP button on the front panel of the unit allows you to
initiate and terminate blood pressure measurements, which are displayed in the NIBP area of
the monitor.
The Propaq MD unit’s NIBP module measures the oscillometric pulses transmitted through the
blood pressure cuff and hose, and calculates the blood pressure measurements accordingly. The
pressure measurement proceeds as follows:
1. The cuff inflates to a preconfigured pressure (determined by patient type) above the
patient’s systolic blood pressure. As the cuff inflates, the Propaq MD unit measures the
oscillometric pulses transmitted from the cuff through the hose, and the SureBP monitoring
software calculates the patient’s systolic, diastolic, and mean blood pressure. Once these
measurements are taken, the cuff deflates. The Propaq MD unit takes these measurements in
about 15 seconds.
The SureBP feature requires the use of a cuff and dual lumen hose.
2.
If high artifact prevents an accurate measurement during cuff inflation, or if you use a cuff
and single lumen hose, the cuff inflates to its target pressure to occlude blood flow through
the arteries in the monitored limb. The cuff deflates incrementally, allowing blood to flow
through the cuff and into the monitored limb. As blood flows through the partially deflated
cuff, it produces pressure oscillations that are transmitted to the Propaq MD unit through the
hose.The Propaq MD unit measures the oscillometric pulses and uses them to calculate the
corresponding systolic, diastolic, and mean blood pressure. This measurement takes about 30
seconds.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...