The NIBP Numeric Display
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
8-3
3.
The NIBP option automatically adjusts the blood pressure measurement procedure in
response to certain error conditions such as:
The NIBP Numeric Display
When NIBP monitoring has been set up and the Propaq MD unit has begun taking NIBP
measurements, the systolic, diastolic, and mean blood pressure measurements appear on the
NIBP numeric display as follows:
The following sections describe how to set up NIBP monitoring.
NIBP Setup and Use
To take safe and accurate NIBP measurements using the Propaq MD unit, you must perform the
following steps, each of which corresponds to a section in this chapter. Read each section
carefully before you perform NIBP measurements.
1. Select the correct size cuff.
2.
Apply the cuff to the patient.
3.
Connect the inflation hose to the Propaq MD unit and to the cuff.
4.
Configure NIBP alarms and settings (if the current NIBP alarms and settings are not
appropriate).
5.
Press the NIBP button on the Propaq MD unit’s front panel and take the blood pressure
measurement.
Condition
Adjustment/Response
The unit cannot detect systolic
pressure
The unit automatically increases the cuff inflation
pressure and completes the blood pressure mea-
surement
The unit cannot detect systolic,
diastolic or mean pressure after
180 seconds
The unit aborts the blood pressure measurement and
deflates the cuff
The unit detects a fault
The unit displays a corresponding error message on
the monitor and aborts the measurement
138
89
(112)
NIBP
mmHg
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...