On the NIBP Parameter Control Panel, you can select an alarm configuration option (
SYS
,
DIA
,
or
MAP
) to enable/disable alarms and set high and low alarm limits (as described previously).
You can also configure these NIBP options:
•
NIBP Mode (
Mode
)
•
NIBP Automatic Measurement Interval (
Auto Interval
)
•
Enable/Disable Smartcuf motion artifact filter
(Smartcuf)
•
Start/Stop Turbocuf Measurements (
Turbocuf
)
Specify NIBP Mode
You can specify that the Propaq MD unit operate in either
Manual
or
Automatic Mode
.
In Manual Mode, the Propaq MD unit takes a single NIBP measurement when you press the
NIBP button on the front panel(
). To repeat the NIBP measurement, you must press the
NIBP key again (select
Manual
).
In Automatic Mode, the Propaq MD unit takes the first of a series of NIBP measurements when
the
Auto Interval
timer expires, and then repeats the NIBP measurement at this specified
interval.
Note:
You can press the NIBP button and manually initiate an NIBP reading at any time
while in Automatic Mode. Manually initiated NIBP readings will not affect the timing
of subsequent NIBP readings in Automatic Mode.
Specify the Automatic NIBP Measurement Interval
You can specify the time interval between NIBP measurements in Automatic Mode. The
default interval between measurements is 5 minutes.
You can specify intervals of
1
,
2
,
3
,
5
,
10
,
15
,
30
,
and
60
minutes between NIBP measurements.
Enable/Disable Smartcuf Motion Artifact Filter
Enabling the use of the Smartcuf motion artifact filter increases the accuracy of NIBP
measurements when motion artifact or weak pulses are present.
NIBP measurements can be adversely affected by many factors, such as cardiac arrthythmias,
sudden changes in blood pressure, body motions such as shivering and convulsions, bumping
the cuff, vibration, vehicle motion, or weak pulses. The Smartcuf filter synchronizes the NIBP
reading with the R-wave of the patient’s ECG to eliminate noise created by patient motion or
vibration.
Note:
The Propaq MD unit must perform ECG monitoring when using Smartcuf.
If artifact is severe, a special symbol appears on the display and printouts:
Figure 8-5 High Artifact Symbol
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...