Applying the SpO
2
Sensor
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
10-7
Applying a Reusable SpO
2
Sensor/Cable
The following instructions describe how to apply a reusable Rainbow sensor. For all other
reusable sensors, refer to the sensor packaging and the manufacturer’s
Directions for Use
for
instructions on how to apply the sensor.
Note:
A reusable sensor is not intended for use on the thumb or across a child’s hand or foot.
After selecting a monitoring site, apply the reusable sensor as follows:
1. Place the selected digit over the sensor window, making sure that the sensor cable runs over
the top of the patient’s hand.
2.
On finger sites, make sure the tip of the finger touches the raised digit stop inside the sensor.
If the fingernail is long, it may extend over and past the finger stop.
3.
Check the sensor position to ensure that the top and bottom halves of the sensor are parallel.
To ensure accurate data, you must have complete coverage of the detector window.
Note:
With smaller digits, the digit may not need to be pushed all the way to the stop to
completely cover the detector window.
4.
Based on the type of patient cable you are using, do one of the following:
•
If you are using a direct one-piece patient cable, see “Connecting the SpO
2
Sensor” on
page 10-8.
•
If you are using the two-piece patient cable, lift the clear plastic protective cover from
the female end of the extension cable, then plug the sensor cable’s male connector into
the extension cable connector -- make sure that the plug is all the way into the connector
(see Figure 10-1).
5.
Lower the clear plastic cover over the connection to secure it (see Figure 10-2).
6.
See “Connecting the SpO
2
Sensor” on page 10-8 to connect the cable to the unit.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...