System Messages
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
10-11
Specifying the SpO
2
Averaging Time
The Masimo SpO
2
module in the Propaq MD unit provides three different time periods over
which SpO
2
values are averaged: 4 seconds, 8 seconds (default) and 16 seconds. The averaging
period is rarely changed from the 8 second default setting. For high risk patients with rapidly
changing SpO
2
conditions, use the 4 second setting. Use the 16 second setting only when the 8
second setting (default) is inadequate due to extremely high artifact conditions.
Selecting the SpO
2
Sensitivity
You can select either normal or high sensitivity for SpO
2
monitoring. The Normal sensitivity
setting is recommended setting for most patients. The High sensitivity setting allows you to
monitor SpO
2
even under very low perfusion conditions. Such conditions may include severe
hypotension or shock. With the High sensitivity setting, however, SpO
2
results are more easily
contaminated by artifact. To ensure accurate SpO
2
readings when using the High sensitivity
setting, observe the patient carefully and continuously.
Selecting the Heart Rate/ Pulse Rate (HR/PR) Tone
The unit allows you enable or disable the tone that the monitor uses to indicate detection of the
patient’s pulse:
On
or
Off
(no tone sounds). The default tone is
Off
.
SpO
2
System Messages
The Propaq MD unit may display the following system messages when monitoring SpO
2
:
System Message
Cause
INITIALIZING
The SpO
2
pulse oximeter is initializing.
SEARCHING
The unit is searching for a pulse.
CHECK SENSOR
The SpO
2
sensor has become disconnected from the
unit, or the sensor is no longer on the patient. Check
sensor and then reconnect it to the unit or reapply it to the
patient.
SENSOR FAILURE
The SpO
2
sensor is defective. Replace the sensor.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...