Functional Testers and Patient Simulators
Some models of commercially available bench top functional testers and patient simulators can
be used to verify the proper functionality of Masimo pulse oximeter sensors, cables and
monitors. See the individual testing device's operator's manual for the procedures specific to the
model of tester that you are using.
While such devices may be useful for verifying that the pulse oximeter sensor, cable, and
monitor are functional, they are incapable of providing the data required to properly evaluate
the accuracy of a system's SpO
2
measurements.
Fully evaluating the accuracy of the SpO
2
measurements requires, at a minimum,
accommodating the wavelength characteristics of the sensor and reproducing the complex
optical interaction of the sensor and the patient's tissue. These capabilities are beyond the scope
of known bench top testers, including known devices which claim to measure sensor LED
wavelength.
SpO
2
measurement accuracy can only be evaluated in vivo by comparing pulse oximeter
readings with values traceable to SpO
2
measurements obtained from simultaneously sampled
arterial blood using a laboratory CO-oximeter.
Many functional testers and patient simulators have been designed to interface with the pulse
oximeter's expected calibration curves and may be suitable for use with Masimo monitors and/
or sensors. Not all such devices, however, are adapted for use with the Masimo digital
calibration system. While this will not affect use of the simulator for verifying system
functionality, displayed SpO
2
measurement values may differ from the setting of the test
device.
For a properly functioning monitor, this difference will be reproducible over time and from
monitor to monitor within the performance specifications of the test device.
LOW PERFUSION
Perfusion has dropped below 20%.
SPO2 DISABLED
A system error has occurred. The Propaq MD unit cannot
take SpO
2
measurements and should be serviced.
System Message
Cause
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...