C
HAPTER
11 M
ONITORING
I
NVASIVE
P
RESSURES
(IBP)
11-4
www.zoll.com
9650-0806-01 Rev. D
Figure 11-2 IBP Channel Parameter Control Panel
6.
Select
Zero Probe
. The unit displays the message
ZEROING
in the IBP Channel’s numeric
display.
7.
The message
ZEROED
appears in the IBP channel’s numeric display.
8.
Close the transducer’s stopcock.
9.
If the unit was unable to zero the transducer, the message
ZERO REJECTED
appears in the
IBP channel’s numeric display. The unit will not display pressure values for the IBP channel
until it has successfully zeroed the transducer and established an acceptable zero reference.
Check that the unit is open to atmospheric air and that it is properly connected to the unit, then
try zeroing the transducer again. The Propaq MD unit will not zero the transducer if it detects
pulsation in the pressure channel, if there is too much noise in the signal, or if transducer’s
offset is too great.
If you cannot zero the transducer after several attempts, replace the transducer or the transducer
cable.
Rezeroing a Transducer
You can re-zero a transducer at any time by opening the transducer stopcock to atmospheric air.
If the unit accepts the new zero reference value, the unit displays pressure values that are based
on that value and adjusts the waveform according to the new scale.
Warning!
If you attempt to zero an IBP channel after the channel has been successfully zeroed
and is currently monitoring a pressure waveform, the unit will display the message,
ZERO REJECTED
, in the IBP channel’s numeric window. This message will preempt
the display of the valid invasive pressure numerics.
P1
Lower
Upper
P1 Systolic Alarm
90
150
Source Label
P1
Display Format
S/D (M)
P1 Diastolic Alarm
60
90
P1 Mean Alarm
70
120
Zero Probe
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...