AED Operation
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
13–5
2 Analyze
WARNING!
Do not analyze the patient ECG during patient movement. A patient must be motionless
during ECG analysis. Do not touch the patient during analysis. Cease all movement via
stretcher or vehicle before analyzing the ECG.
The Propaq MD automatically begins the analysis of the patient’s ECG rhythm,
displays an
ANALYZING ECG message for 5 seconds, and announces and displays a
STAND CLEAR
message. If therapy electrodes have not been properly connected to the patient, an
ATTACH
PADS
or
CHECK PADS
message is displayed and analysis will be inhibited.
Note:
You can press
Pause
to suspend the rescue cycle. While the rescue cycle is paused,
ECG analysis continues in the background. The trends, log, and 12-lead features are
only accessible when the rescue cycle is suspended. Press the
Analyze
quick access
key to resume the analysis process.
Note:
If the Propaq MD has been configured to perform extra CPR at startup, it displays a
CHECK PULSE
voice prompt and the message appears for 10 seconds. Then it
displays a
PERFORM CPR
message along with a voice prompt for the configured
duration before analysis begins. You can start an ECG analysis during the CPR interval
by pressing
Analyze
.
An ANALYZING ECG message is displayed while the patient’s ECG is analyzed. Once the
analysis is completed, the unit indicates whether or not a shock is advised.
WARNING!
ECG rhythm analysis does not warn of patient asystole, which is not a shockable
rhythm.
STAND
CLEAR
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...