C
HAPTER
13
A
UTOMATED
E
XTERNAL
D
EFIBRILLATOR
(AED) O
PERATION
13–8
www.zoll.com
9650-0806-01 Rev. D
Continue Patient Care
Continue patient care according to medical protocols.
Operating Messages
The unit uses both audio and visual prompts to present critical information to operators. The
following information describes the unit default configuration. If your device has been custom
configured, some of the information may be different.
There are 10 voice prompts used in AED mode. Most of these prompts are accompanied by a
message displayed on the monitor. The voice prompts are given only once, but the monitor
continues to display the message until new action is taken by the operator or the device status
changes.
The unit will alternately display two different messages in the same field of the display when
two conditions are detected at the same time. For example, a
LOW BATTERY
message may
alternately display on the same line of the monitor as the
CHECK PADS
message.
Audio and Display Messages
The display messages and voice prompts that can occur during AED operation are described
below.
ATTACH PADS
If the unit is powered on without therapy pads connected to the patient, the
ATTACH PADS
message is announced and displayed.
ANALYZING ECG/STAND CLEAR
The
ANALYZING ECG
message is displayed and the
STAND CLEAR
message is displayed and
announced when the ECG analysis starts automatically or after pressing the
Analyze
quick
access key. They indicate that an active ECG analysis is in progress.
SHOCK ADVISED
A shockable rhythm has been detected and defibrillation is advised. The selected energy level is
displayed.
PRESS SHOCK
This message is displayed and announced when the ECG analysis has determined that a shock
is advised and the selected energy is ready to be delivered.
SHOCKS: XX
This message displays the number of shocks that have been delivered by the unit since it was
powered on. It resets to 0 after the unit has been off for more than two minutes.
NO SHOCK ADVISED
When ECG analysis detects a nonshockable rhythm, this message is announced and displays
for 10 seconds following completion of the analysis.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...