the use of high quality silver/silver chloride (Ag/AgCl) electrodes to minimize this
effect; the circuitry in the instrument returns the trace to the monitor display within a
few seconds.
•
Wait 15 seconds after defibrillator discharge before attempting a 12-lead acquisition.
Electrode polarization subsequent to defibrillator discharge may result in excessive
noise on the 12-lead ECG printout.
•
When not in use, cover the patient cable's V-lead connector with the supplied plastic
cap. Failure to do so may result in a shock hazard during defibrillation attempts.
•
To assure protection against the effects of defibrillator discharge, use only 12-lead
cables supplied by ZOLL Medical Corporation.
•
Check the operation and integrity of the Propaq MD unit and 12-lead cable regularly
by performing the Daily Operational Verification Test.
•
When attempting to interpret subtle ECG changes (such as ST segments) use only the
Diagnostic frequency response setting. Other frequency response settings may cause
misinterpretation of the patient’s ECG.
•
Use only ZOLL-approved accessories to ensure the Type CF defibrillator proof rating
of the 12-Lead input.
•
Implanted pacemakers may cause the heart rate meter to count the pacemaker rate
during incidents of cardiac arrest or other arrhythmias. Carefully observe pacemaker
patients. Check the patient's pulse; do not rely solely on heart rate meters. Dedicated
pacemaker detection circuitry may not detect all implanted pacemaker spikes. Patient
history and physical exam are important in determining the presence of an implanted
pacemaker.
Entering Patient Information
To enter patient information, press the 12-Lead quick access key (
), then the Patient Info
quick access key (
). The screen displays the Patient Info parameter panel, in which you can
enter the patients’s name, age, gender, and identification number: The screen displays the
Patient Info parameter panel, in which you can enter the patients’s name, age, gender, and
identification number:
12
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...