12-Lead Print and Display Options
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
14-13
or a 3 x 4 Cabrera format:
The Propaq MD unit stores a minimum of 32 12-Lead snapshots in a separate log. Once 32 12-
lead snapshots are stored, the oldest snapshot in the log is overwritten by subsequent snapshots.
Note that if a patient data log is full, you will not be able to store12-lead snapshots. The 12-lead
snapshots are erased when you clear the log.
When you are done viewing and printing the 12-Lead waveform traces, press the Exit 12 quick
access key (
)to restore the display of other monitoring functions.
12-Lead Print and Display Options
The Propaq MD unit provides additional 12-Lead print and display options that you can specify
through the Supervisor parameter control panels (access to Supervisor is passcode-controlled).
Press the Setup quick access key (
), then select Supervisor
.
Using the navigation keys,
select the four digits in the Supervisor passcode. Press
SAVE
when you are finished. Once you
have entered your supervisor passcode, you will be able to access the configurable options in
the Supervisor menu.
Select the Supervisor> ECG> 12-Lead option to display the 12-Lead parameter control panel:
12
Exit
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...