C
HAPTER
18
E
XTERNAL
P
ACING
18–6
www.zoll.com
9650-0806-01 Rev. D
Pacing in Fixed Mode
If ECG electrodes are not available or there is some circumstance that prevents or interferes
with the surface ECG, the Propaq MD delivers pacemaker pulses at a fixed rate.
Fixed pacing should be performed only in an emergency when no alternative is available.
1 Turn On Unit
Press the green power switch located on the top of the unit. The green, yellow, and red lights at
the top of the unit flash on and off, and the unit displays the message SELF TEST PASSED.
2 Apply ECG Electrodes/Hands-Free Therapy Electrodes
Apply ECG electrodes, attach lead wires, and connect the ECG cable to the Propaq MD side
panel (see Chapter 6 for instructions on attaching ECG electrodes to the patient). Attach
hands-free therapy electrodes according to instructions on the electrode packaging. Connect
these therapy electrodes to the MFC.
Therapy Electrode Application
WARNING!
Poor adherence and/or air under the therapy electrodes can lead to the possibility of
arcing and skin burns.
1. Apply one edge of the pad securely to the patient.
2.
Roll the pad smoothly from the applied edge to the other, being careful not to trap any air
pockets between the gel and skin.
3.
Ensure that hands-free therapy electrodes are making good contact with the patient’s skin and
are not covering any part of any other ECG electrodes.
4.
Press the Lead quick access key and select I, II, or III to provide the largest amplitude QRS
complex.
Note:
When the Pacer is on, the lead selection is restricted to Leads I, II, or III.
5.
Verify that R-waves are being properly detected by confirming that a QRS tone occurs with
each displayed R wave or by verifying that the Propaq MD unit’s heart rate display accurately
reflects the patient’s pulse rate.
Pad
Skin
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...