CPR Voice Prompts
The Propaq MD unit can be configured to issue voice prompts related to the depth of chest
compressions as feedback to rescuers performing CPR. Two voice prompts are available for
this purpose:
•
Push Harder
•
Good Compressions
When chest compressions are detected but their depth is consistently less than the target depth
(AHA/ERC 2005: 1.75, AHA/ERC 2010: 2.0), the device periodically issues the “Push Harder”
voice prompt. If the rescuer responds by increasing compression depth to more than the target
depth on a consistent basis, the unit issues a “Good Compressions” prompt.
CPR Metronome
The Propaq MD unit includes a CPR metronome feature that can be used to encourage rescuers
to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per
minute.
The CPR metronome can be configured to operate in all modes (AED, Manual, and Advisory/
CPR Protocol), or in AED mode only. The CPR metronome can also be disabled for all modes.
When activated, the metronome beeps at the AHA/ERC recommended rate to provide a
compression rhythm for rescuers to follow.
The CPR metronome’s Constant Metronome feature allows the metronome to continue beeping
under certain circumstances. When Enable Metronome is set to Never in the Supervisor>CPR
menu, Constant Metronome is greyed out.
The CPR metronome can be configured as follows.
Constant Metronome Enabled
AED Mode
•
If Enable Metronome is set to Always or AED Mode Only, the metronome is active when
chest compressions are first detected during the CPR interval. The metronome is silent when
the CPR idle timer starts.
•
If set to Never, the metronome is silent.
Advisory/CPR Protocol or Manual Mode
•
If Enable Metronome is set to Always, the metronome is active when chest compressions
are first detected during the CPR interval. The metronome is silent when the CPR idle timer
starts.
•
If set to AED Mode Only or Never, the metronome is silent.
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...