CPR Dashboard
9650-0806-01 Rev. D
Propaq MD Operator’s Guide
19–3
Constant Metronome Disabled
AED Mode
•
If Enable Metronome is set to Always or AED Mode Only, the metronome is active when
chest compressions are first detected during the CPR interval.
•
If set to Never, the metronome is silent.
Advisory/CPR Protocol or Manual Mode
•
If Enable Metronome is set to Always, the metronome is active only when the reported
compression rate falls below 80 cpm during the CPR period. If the chest compression sensor
is connected, the metronome is suspended after compressions have not been detected for at
least 2 seconds.
•
If set to Never or AED Mode Only, the metronome is silent.
The following table indicates when the CPR metronome is active.
CPR Dashboard
Whenever CPR-D-padz or OneStep electrodes
are connected to the Propaq MD defibrillator
and it senses compression, the unit illuminates the CPR Dashboard in the lower center portion
of the display. This field includes the CPR feedback indicators described in the following
sections.
Enable
Metronome
Setting
AED Mode
Advisory/CPR
Protocol Mode
Manual Mode
Constant Metronome Enabled
AED Mode
Only
On when CPR detected
Off when CPR Idle
Off
Off
Always
On when CPR detected
Off when CPR Idle
On when CPR detected
Off when CPR Idle
On when CPR detected
Off when CPR Idle
Never
Off
Off
Off
Constant Metronome Disabled
AED Mode
Only
On at start of CPR period
Off
Off
Always
On at start of CPR period
On only at <80
compressions per minute
On only at <80
compressions per minute
Never
Off
Off
Off
CPR Release Indicator
Chest Compression
Indicator
Summary of Contents for Propaq MD
Page 1: ...Propaq MD Operator s Guide 9650 0806 01 Rev D...
Page 40: ...CHAPTER 1 GENERAL INFORMATION 1 28 www zoll com 9650 0806 01 Rev D...
Page 60: ...CHAPTER 2 PRODUCT OVERVIEW 2 20 www zoll com 9650 0806 01 Rev D...
Page 74: ...CHAPTER 4 TRENDS 4 4 www zoll com 9650 0806 01 Rev D...
Page 84: ...CHAPTER 5 ALARMS 5 10 www zoll com 9650 0806 01 Rev D...
Page 94: ...CHAPTER 6 MONITORING ECG 6 10 www zoll com 9650 0806 01 Rev D...
Page 104: ...Chapter 7 Monitoring Respiration Resp and Heart Rate HR 7 10 www zoll com 9650 0806 01 Rev D...
Page 120: ...CHAPTER 8 MONITORING NON INVASIVE BLOOD PRESSURE NIBP 8 16 www zoll com 9650 0806 01 Rev D...
Page 154: ...CHAPTER 11 MONITORING INVASIVE PRESSURES IBP 11 10 www zoll com 9650 0806 01 Rev D...
Page 160: ...CHAPTER 12 MONITORING TEMPERATURE 12 6 www zoll com 9650 0806 01 Rev D...
Page 186: ...CHAPTER 14 12 LEAD ECG INTERPRETIVE ANALYSIS 14 16 www zoll com 9650 0806 01 Rev D...
Page 212: ...CHAPTER 17 ADVISORY CPR PROTOCOL DEFIBRILLATION 17 6 www zoll com 9650 0806 01 Rev D...
Page 222: ...CHAPTER 18 EXTERNAL PACING 18 10 www zoll com 9650 0806 01 Rev D...
Page 240: ...CHAPTER 21 PATIENT DATA 21 6 www zoll com 9650 0806 01 Rev D...
Page 268: ...CHAPTER 22 COMMUNICATIONS 22 28 www zoll com 9650 0806 01 Rev D...
Page 284: ...CHAPTER 24 MAINTENANCE 24 12 www zoll com 9650 0806 01 Rev D...
Page 326: ...ACCESSORIES B 6 www zoll com 9650 0806 01 Rev D...